The LATITUDE Communicator is suitable for use in a professional health care
facility or home environment.
Test
Compliance
Electromagnetic
environment—guidance
Conducted disturbances induced
by RF fields
3 V/m from 0.15 MHz to 80 MHz
6 V/m in ISM bands from 0.15
MHz to 80 MHz
The ISM bands between 0.15 MHz
and 80 MHz are 6.765 MHz to
6.795 MHz 13.553 MHz to 13.567
MHz 26.957 MHz to 27.283 MHz
40.66 MHz to 40.70 MHz.
The amateur radio bands
between 0.15 MHz and 80
MHz are
1.8 MHz to 2.0 MHz
3.5 MHz to 4.0 MHz
5.3 MHz to 5.4 MHz
7.0 MHz to 7.3 MHz
10.1 MHz to 10.15 MHz
14.0 MHz to 14.2 MHz
18.07 MHz to 18.17 MHz
21.0 MHz to 21.4 MHz
24.89 MHz to 24.99 MHz
28.0 MHz to 29.7 MHz
50.0 MHz to 54.0 MHz.
Voltage dips
a
0% U
T
for 0.5 cycle at 0°,
45°, 90°, 135°, 180°,
225°, 270°, and 315°
0% U
T
for 1 cycle and
70% U
T
for 25/30 cycles at 0°
Voltage interruptions
a
0% U
T
for 250/300 cycles
a. Voltage dips and interruptions: UT is the AC mains voltage prior to application of the test level.