10.14 Replacing the pump
10.15 Checking the pressure sensor
10.16 Checking the response time of
the pressure sensor
i
Please also refer to Sections 12.1 and 12.5 dealing with the electrical wiring and
hosing for the EAM exhaust-gas analyzer module.
1.
Remove the 2-core line leading from the pump to the control module PCB.
2.
Remove the hose lines from the pump heads.
3.
Using a diagonal cutter, cut the cable binder (on the fastening clip) and remove.
4.
Pull the pump out of the fastening clip.
5.
Press the new pump into the fastening clip.
6.
Insert the 2-core line leading from the pump in the control module PCB.
7.
Push the hose lines onto the hose connectors of the pump as shown in the gas circuit
diagram (Sec. 12.5) and fasten with hose clips.
8.
Perform a leak test (Sec. 4.2) and function test (Sec. 10.13).
!
As from FD 07.2004 pumps of a shorter design (1 687 222 167) will be installed. In
order to fasten these in place, angle bracket 1 681 332 269 also has to be fitted. For
mounting details see the Installation instructions 1 689 978 483.
1.
Start the leakage test (Sec. 4.2)
using the Service software.
If a leak test can now be performed
and it is passed, the pressure sensor
is deemed to be in order.
If a leakage test cannot be perfor-
med, a message appears prompting
you to remove the seal.
In this case, check the voltage supply
to the pressure sensor (Secs. 10.2
and 10.3).
If the voltage supply is in order, the
pressure sensor must be replaced
(Sec. 10.17).
1.
Using the Service software, start the
Read measurement values
menu.
2.
Following zero calibration and the
HC residues test, seal off the test-gas
inlet.
3.
After 5 - 7 seconds, the message
In-
sufficient flow
should be displayed.
If the above message is not displayed,
you must perform a leak test (Sec. 4.2).
If the leak test is failed, check that the
hosing conforms to the gas circuit
diagram (Sec. 12.2).
If the hosing is in order, replace the
pressure sensor (10.17).
Содержание BEA 150
Страница 91: ...92 12 Overview 12 1 Electrical wiring Siemens analysis chamber 459716 2Ha ...
Страница 94: ...95 12 4 Hosing Siemens analysis chamber X26 X24 X25 X22 X5 LP 223 X19 459716Ht ...
Страница 96: ...97 12 6 Electrical wiring Andros analysis chamber 4597162Ht ...
Страница 99: ...100 12 9 Hosing Andros analysis chamber 459869_03Ht ...
Страница 100: ...101 12 10 Gas circuit diagram Andros analysis chamber ...