18
4.7.1
Pump On
4.7.2
Pump Off
4.7.3
Install new O2 sensor
4.7.4
O
2
measurement
Select this menu by double-clicking the
mouse or with the
o u
keys and confirm
with
E
.
The pump is switched on and the
OK
box
is displayed for 1 s.
This function can be used, for example, to
purge the exhaust-gas analyzer module if
it has been contaminated with HC.
Select this menu by double-clicking the
mouse or with the
o u
keys and confirm
with
E
.
The pump is switched off and the
OK
box
is displayed for 1 s.
See Section 10.11.
Start the menu by double-clicking the
mouse or with
E
.
The pump is switched on and zero calibra-
tion takes place. During this process, the
O
2
measuring channel is recalibrated to
20.9% O
2
(this same calibration is also
performed automatically during each
system calibration with zero gas/air).
After zero calibration, the pump is swit-
ched off.
The exhaust-gas analyzer module displays
the concentration of oxygen with the cur-
rent ADC voltage value below it.
If the current ADC voltage lies at the lower
threshold of 7 mV, you must replace the
O
2
measuring sensor (Section 10.11).
Nominal voltage:
7 mV - 11 mV
Содержание BEA 150
Страница 91: ...92 12 Overview 12 1 Electrical wiring Siemens analysis chamber 459716 2Ha ...
Страница 94: ...95 12 4 Hosing Siemens analysis chamber X26 X24 X25 X22 X5 LP 223 X19 459716Ht ...
Страница 96: ...97 12 6 Electrical wiring Andros analysis chamber 4597162Ht ...
Страница 99: ...100 12 9 Hosing Andros analysis chamber 459869_03Ht ...
Страница 100: ...101 12 10 Gas circuit diagram Andros analysis chamber ...