8.13.3 Error messages for "LBCTEST"
No.
Possible cause
Remedy
Chapter
$0001
Bus+ fault or
Repeat test using short circuit plug KS 350.
$0002
OBD/CAN-LP defective
If malfunction no longer present, OBD/CAN-
$0003
connecting cable should be replaced.
If malfunction still present, OBD/CAN-LP
to be replaced.
10.30
$0004
Bus fault or
Repeat test using short circuit plug KS 350.
$0008
OBD/CAN-LP defective
If malfunction no longer present, OBD/CAN-
$0012
connecting cable should be replaced.
If malfunction still present, OBD/CAN-LP
to be replaced.
10.30
8.13.4 Error messages for "CANTEST"
No.
Possible cause
Remedy
Chapter
$0001
OBD/CAN connection cable CAN-L
Repeat test using short circuit plug KS 350.
$0004
or OBD/CAN-LP defective.
If malfunction no longer present, OBD/CAN-
connecting cable should be replaced.
If malfunction still present, OBD/CAN-LP
to be replaced.
10.30
$0002
OBD/CAN connection cable CAN-H
Repeat test using short circuit plug KS 350.
$0008
or OBD/CAN-LP defective.
If malfunction no longer present, OBD/CAN-
connecting cable should be replaced.
If malfunction still present, IOBD/CAN-LP
to be replaced.
10.30
Содержание BEA 150
Страница 91: ...92 12 Overview 12 1 Electrical wiring Siemens analysis chamber 459716 2Ha ...
Страница 94: ...95 12 4 Hosing Siemens analysis chamber X26 X24 X25 X22 X5 LP 223 X19 459716Ht ...
Страница 96: ...97 12 6 Electrical wiring Andros analysis chamber 4597162Ht ...
Страница 99: ...100 12 9 Hosing Andros analysis chamber 459869_03Ht ...
Страница 100: ...101 12 10 Gas circuit diagram Andros analysis chamber ...