Blue-White ProSeries SONIC-PRO S4 Series Скачать руководство пользователя страница 9

Page 7

Sonic-Pro

5.1
Power On

5.2
Run Screen

5.0 Operation

1. If power is present on Terminal T1 pins 1 & 2,  the meter is always powered on.
 
2. If power is removed from Terminal T1 pins 1 & 2, and 4 D-Cell batteries are installed on 
Terminal T1 pins 4 & 5, the meter will stay on (battery back-up). Under these conditions, the 
meter can be turned on/off by pressing and holding the front panel ENTER button for three 
seconds.

3. If power is 

only 

present on the battery Terminal T1 pins 4 & 5, then the meter can be 

turned on/off by pressing and holding the front panel ENTER button for three seconds. 
Operating time is limited when powered by batteries only.

When the S4 is factory ordered with a pipe fitting/transducer, the pipe fitting data and the 
default display feature parameters are pre-configured. When power is applied and there is 
water in the pipe fitting, the following RUN display screen will be displayed.

Error Code

There are three possible FAULT indicator screens.

Ÿ

FAULT 11 - 

Transducers are not connected or there is no fluid in the pipe.

Ÿ

FAULT 12 - 

The allowable SOS (Speed Of Sound) error set-point has been exceeded.

Ÿ

FAULT 21 - 

The configuration data is not valid.

Flow Direction

Total Units

U.S Gallons/Min

105

1482

25

05

469

U.S Gallons

.

.

0.28”

Transducer Separation

Distance

Flow Rate Units

Flow Total

Flow rate

Measured Sound Speed

Battery Life

Measurement Reliability

U.S Gallons/Min

F-11

0”

0

U.S Gallons

0.28”

XDCR Disc.

(press and hold ESC to clear)

Page 8

Sonic-Pro

5.3
Menu
Navigation

The Configuration Menu

From the RUN mode, press and release the 

ENTER

 button to enter the CONFIGURATION 

menu. The following menu will be displayed:

Paddle Wheel

XDCR Settings

Meter Settings

Output Settings

Clock

Calibration

Global Config.

About

5.3.1
Selecting and 
Entering Menus

Ÿ

Press up arrow 

p

 or down arrow 

q

 to select menu items. 

The selected item will be highlighted.

Ÿ

Press right arrow 

u

 or 

ENTER 

to enter the selected menu.

Ÿ

Press left arrow 

t

 to go back one menu.

Ÿ

Notice the gray area. These menu items are hidden from 
view. Press the up arrow 

p

 or down arrow 

q

 to display 

hidden menu items. 

5.3.3
Inputting Data

Ÿ

Press right arrow 

u

 or left arrow 

t

  to select a digit to 

modify. Pressing the left arrow 

t

 will create additional 

numbers to the left.

Ÿ

Press up arrow 

p

 to increase or down arrow 

q

 to 

decrease the value.

Ÿ

Press 

ENTER

 to confirm the modifications and return to 

the menu heading. 

Ÿ

Press 

ESC

 to cancel the modifications and return to the 

menu heading.

5.3.2
Selecting Optional
Items

Ÿ

Press up arrow 

p

 or down arrow 

q

 to select optional 

configuration items. The selected item will be highlighted.

Ÿ

Press left arrow 

t

 or 

ENTER 

to choose the selected 

optional item and return to the menu heading.

U.S. Barrels Oil

Cubic Feet

Acre Feet

Imperial Gallons

Cubic Meters

Liters

Milliliters

FR Custom V/Gal

U.S. Gallons

Ounces

U.S. Barrels Liq.

Enter Value

9.1000

0

5.3.4
Save Configuration

Ÿ

Press left arrow 

t

  to exit to the main menu.

Ÿ

Press down arrow 

q

 to Highlight 

Save Config

.

Ÿ

Press right arrow 

u

 or 

ENTER 

to save the configuration 

changes. 

Ÿ

Press 

ENTER 

to confirm the configuration changes have 

been saved.

Ÿ

Press up arrow 

p

 to RUN. 

Ÿ

Press right arrow 

u

 or 

ENTER 

to enter the RUN mode,

Configuration

XDCR Settings

Save Config.

Configuration

Configuration Saved!

OK

XDCR Settings

Save Config.

Configuration

5.1.1
Battery Powered
operation

The following meter features are disabled during battery operation:

Ÿ

Communications Module 

Ÿ

Analog output

Ÿ

Pulse Frequency Output

Continuous Running

On / Sleep Time (seconds)

10 / 10

128 Hr

72 Hr

310 Hr

377 Hr

450 Hr

10 / 100

5 / 100

3 / 268

Ÿ

Press the 

ENTER

 button to enter the menu system. 

Ÿ

Press the 

ESC 

button to clear the flow total to zero (if enabled in configuration).

Ÿ

Press the up arrow 

p

 to toggle the flow rate to linear Feet/Sec., Meters/sec and back to 

the configured user units of measure.

5.2.1
Fault Screen

While the batteries pack is provided primarily as back-up power, the S4 meter can be 
operated solely with 4 D-cell batteries (not included). The meter can be configured to sleep 
between measurements in order to conserve power. The approximate battery life is .....

Industries, Ltd.

Industries, Ltd.

ProSeries

by Blue-White Ind.

d

Industries, Ltd.

Industries, Ltd.

ProSeries

by Blue-White Ind.

d

Содержание ProSeries SONIC-PRO S4 Series

Страница 1: ...Installation and Operating Manual 2 TWO YEAR WARRANTY ProSeries Series S4 Ultrasonic Flow Meter Industries Ltd Industries Ltd ProSeries by Blue White Ind d SONIC PRO d...

Страница 2: ...DING THE SAFE USE OF THIS PRODUCT AND OTHER TECHNICAL ASSISTANCE MAY BE DIRECTED TO Blue White Industries 714 893 8529 techsupport blue white com CAUTION risk of electric shock CAUTION risk of danger...

Страница 3: ...ions of other chemicals or other chemicals entirely may be acceptable provided the chemical does not change the sound speed SOS more than approximately 100 meters per second and provided the fluid all...

Страница 4: ...must not be cut or modified Note that the Sonic Pro can accurately measure flow from either direction The SPU can be mounted on a wall or on a horizontal or vertical run of pipe Although the Sonic Pro...

Страница 5: ...is required on downward flows to ensure a full pipe See the minimum straight length of pipe requirement chart above Related piping must be independently supported The meter should not be allowed to s...

Страница 6: ...d tight cable gland connectors for cable diameter from 200 to 394 inches 5 1 to 10 0 mm Two liquid tight cable gland connectors cable for diameters from 118 to 255 inches 3 0 to 6 5 mm One communicati...

Страница 7: ...ector T2 1 2 Positive Negative Type Open collector w internal 1 K pull up to 5V Can be pulled up to 30V 10mA 3K pull up Duty Cycle 50 5 Frequency Range 0 1000 Hz Resolution 1 Hz Accuracy 0 5 F S PADDL...

Страница 8: ...allowable SOS Speed Of Sound error set point has been exceeded FAULT 21 The configuration data is not valid Flow Direction Total Units U S Gallons Min 105 1482 25 05 469 U S Gallons 0 28 Transducer S...

Страница 9: ...he value Press ENTER to confirm the modifications and return to the menu heading Press ESC to cancel the modifications and return to the menu heading 5 3 2 Selecting Optional Items Press up arrow p or...

Страница 10: ...use your inlet outlet pipe material 10 Fluid Type Select WATER or CUSTOM 11 Fluid SOS If WATER is selected it is not necessary to input a fluid SOS If CUSTOM fluid type is selected the Speed of Sound...

Страница 11: ...e factory set inside the pipe fitting and cannot be moved Select the mount method that is shown on the Calibration Detail label affixed to the pipe fitting Transducer Model I C J Mount Method W V Cabl...

Страница 12: ...ings Direction Zero A B Scaling A B Enter Value Sign A B Zero Value A B Positive B A A B Enter Value Zero B A Scaling B A Enter Value Sign B A Zero Value B A Enter Value 0 975 0 455 Negative Positive...

Страница 13: ...n the acceptable range and there are not too many particles or other sound blocking or reflecting contaminants in the fluid which can inhibit the sound beam The target Fluid SOS value can be adjusted...

Страница 14: ...els Oil Cubic Feet Acre Feet Imperial Gallons Cubic Meters Liters Milliliters FR Custom V Gal Fl Total Decimal 0 Enter Value Flow Average s 0 Enter Value Enable Tot Clear Yes No Low Fl Cut Off 0 Enter...

Страница 15: ...into any Tot Set Point 1 2 3 4 or 5 a new log will be generated when the total accumulated flow value is equal to the configured value To access the log entries you must copy the logs onto a USB flash...

Страница 16: ...ters Milliliters TOT Custom V Gal Fl Total Decimal 0 Enter Value Flow Average s 0 Enter Value Enable Tot Clear Yes No Low Fl Cut Off 0 Enter Value High Fl Cut Off 999 Enter Value Scaling Off set 1 000...

Страница 17: ...Off value totalizing will still occur 6 3 2 6 High Flow Rate Cut off Setting High Fl Cut Off 999 Enter Value Scaling Off Set is a multiplier used to correct for flow inaccuracies All flow rate values...

Страница 18: ...e meter was sleeping using the following formula Total flow display value T1 F1 F2 2 x N 1 Where T1 Last total before sleep F1 Last flow rate units per second before sleep F2 Flow rate average of the...

Страница 19: ...gh Release A 0 Enter Value Low Trigger A 0 Enter Value Low Release A 0 Enter Value Time s A 0 Enter Value Monitor Totalizer Batch Amt A 0 Enter Value Contact Timer s A 0 Enter Value Faults A Relay B M...

Страница 20: ...ue Rate at Max Curr 0 0000 Enter Value Max Analog O P 20 Enter Value 6 4 2 Pulse 0 1000 Hz Output Menu A digital pulse output signal can be programmed within the range of 0 1000 Hz pulses per second S...

Страница 21: ...ue Faults B None A None B Relay B Monitor Flow Rate High Trigger B 0 Enter Value High Release B 0 Enter Value Low Trigger B 0 Enter Value Low Release B 0 Enter Value Time s B 0 Enter Value None B 6 4...

Страница 22: ...Value Contact Timer s B 0 Enter Value 6 4 3 2 CC Assigned to Monitor FAULTS When in the Relay A or Relay B menu selecting and then exiting the FAULTS menu will assign that relay to energize when any...

Страница 23: ...unsigned char 8 Frequency output 45 Firmware Version unsigned char 8 8 ASCII characters A11 22 33 53 Log Entry unsigned char 86 Log Entry String Eg 2013 12 29 09 15 00 FR 0 0000 TL 0 00 Description P...

Страница 24: ...e meter can be re calibrated in the field by selecting Calibration and pressing the right arrow u or ENTER button The pipe must be full of fluid and there must be no movement of the fluid in the pipe...

Страница 25: ...Save Config function is possible at the main menu Global Config Password 1234 Enter Value Change Password Change Password Old Pass 1234 OK CANCEL NEW Pass 5678 OK Enter master password 3408 to clear a...

Страница 26: ...me Units Independently configurable Rate and Total display units in U S Gallons ounces barrels US liquid barrels US oil cubic ft acre ft Imperial British gallons liter cubic meter or user defined cust...

Страница 27: ...ipe fitting transducer no transducer cables Standard 4 20mA and High Speed Pulse output only Without power supply customer supplies 15 30 VDC 0 25 ft cable length with field replaceable liquid tight c...

Страница 28: ...03 5 41 1427 10 50 1447 20 68 1481 30 86 1507 40 104 1526 50 122 1541 60 140 1552 70 158 1555 80 176 1555 90 194 1550 100 212 1543 Water Sound Speeds 7 2 Sound speed data Page 27 Sonic Pro Industries...

Страница 29: ...in 18 39cm 5 90in 14 97cm 74in 1 88cm 2 75in 6 99cm 7 24in 18 39cm 5 83in 14 80cm 55in 1 39cm 2 75in 6 99cm Sonic Pro Inline Pipe Fittings Pipe Size 2 1 5 1 75 5 Pipe Size OL OW A B C 2 7 47in 18 97cm...

Страница 30: ...ers TOT Custom V Gal Fl Total Decimal 0 Enter Value Flow Average s 0 Enter Value Enable Tot Clear Yes No Low Fl Cut Off 0 Enter Value High Fl Cut Off 999 Enter Value Scaling Off set 1 0000 Enter Value...

Страница 31: ...Enter Value Time s B 0 Enter Value Monitor Totalizer Batch Amt B 0 Enter Value Contact Timer s B 0 Enter Value Faults B Clock CLOCK SET 2013 07 24 14 31 OK CANCEL Calibration Clear Cal Value None A N...

Страница 32: ...kmanship for 24 months from date of factory shipment WARRANTY COVERAGE IS LIMITED TO REPAIR OR REPLACEMENT OF THE DEFECTIVE FLOWMETER ONLY UNDER NO CIRCUMSTANCES SHALL BLUE WHITE BE LIABLE FOR ANY CON...

Отзывы: