Sonic-Pro

6.3.2
Display
Menu

In the 

DISPLAY

 menu, the following features may be configured:

!

Flow Rate - Volume Units

!

Flow Rate - Time Units

!

Flow Rate - Decimal Point Location

!

Totalizer - Volume Units

!

Totalizer - Decimal Point Location

!

Flow Rate Averaging

!

Enable/Disable clearing the totalizer from the front panel

!

Low Flow Rate Cut-Off

!

High Flow Rate Cut-Off

!

Scaling Offset

!

Zero Offset

!

SOS Fault Window

!

Sleep Time

Display

Flow Rate

Volume Units

Ounces

U.S. Gallons

U.S. Barrels Liq.

U.S. Barrels Oil

Cubic Feet

Acre Feet

Imperial Gallons

Cubic Meters

Liters

Milliliters

FR Custom V/Gal

Time Units

Seconds

Hours

Days

Minutes

Fl. Rate Decimal

0

Enter Value

Totalizer

Volume Units

Ounces

U.S. Gallons

U.S. Barrels Liq.

U.S. Barrels Oil

Cubic Feet

Acre Feet

Imperial Gallons

Cubic Meters

Liters

Milliliters

TOT Custom V/Gal

Fl. Total Decimal

0

Enter Value

Flow Average(s)

0

Enter Value

Enable Tot. Clear?

Yes

No

Low Fl. Cut-Off

0

Enter Value

High Fl. Cut-Off

999

Enter Value

Scaling Off-set

1.0000

Enter Value

Zero Off-set

Sign

FR Offset

0.0000

Enter Value

Negative

Positive

SOS Fault Window

10

Enter Value

Sleep Time(s)

 0

Enter Value

In the 

Flow Rate 

menu

the 

Volume Units 

and 

Time Units 

may be selected from a list. 

A custom volume unit may be created by selecting 

FR Custom V/Gal 

and entering the 

number of gallons in the custom volume.

The 

Fl. Rate Decimal 

decimal point location may be set to a maximum of 5 digits.

6.3.2.1
Flow Rate Menu

Flow Rate

Volume Units

Ounces

U.S. Gallons

U.S. Barrels Liq.

U.S. Barrels Oil

Cubic Feet

Acre Feet

Imperial Gallons

Cubic Meters

Liters

Milliliters

FR Custom V/Gal

Time Units

Seconds

Hours

Days

Minutes

Fl. Rate Decimal

0

Enter Value

0

Enter Value

0

Enter Value

0

Enter Value

Page 16

Sonic-Pro

In the 

Totalizer 

menu, the 

Volume Units 

may be selected from a list. 

A custom volume unit may be created by selecting 

FR Custom V/Gal 

and entering the 

number of gallons in the custom volume.

The 

Fl. Total Decimal 

decimal point location may be set to a maximum of 5 digits.

6.3.2.2
Totalizer Menu

Totalizer

Volume Units

Ounces

U.S. Gallons

U.S. Barrels Liq.

U.S. Barrels Oil

Cubic Feet

Acre Feet

Imperial Gallons

Cubic Meters

Liters

Milliliters

TOT Custom V/Gal

Fl. Total Decimal

0

Enter Value

0

Enter Value

Flow Average(s) 

sets the running average period in seconds. Any number of seconds from 2 

to 10 seconds is possible. 

6.3.2.3
Flow Average
Setting

Flow Average(s)

0

Enter Value

Enable Tot. Clear? 

enables or disables the ability of the user to clear the accumulated total 

flow by pressing the ESC button.

6.3.2.4
Clear Total
Setting

Enable Tot. Clear?

Yes

No

Low Fl. Cut-Off 

sets the minimum displayed flow rate. Calculated flow rate values less than 

this value will be  displayed as Zero and flow totalizing will not occur.

6.3.2.5
Low Flow Rate
Cut-off Setting

Low Fl. Cut-Off

0

Enter Value

High Fl. Cut-Off 

sets the maximum displayed flow rate. Calculated flow rate values greater 

than this value will be displayed as the 

High Fl. Cut-Off value

, totalizing will still occur. 

6.3.2.6
High Flow Rate
Cut-off Setting

High Fl. Cut-Off

999

Enter Value

Scaling Off-Set 

is a multiplier used to correct for flow inaccuracies. All flow rate values will 

be multiplied by the 

Scaling Off-Set 

value and the result displayed and totaled.

6.3.2.7
Scaling Off-Set
Setting

Scaling Off-set

1.0000

Enter Value

Zero Off-Set 

is a factory calibration value used to offset the zero flow setpoint during the 

initial calibration of the 

Ultrasonic Inline Pipe Fitting

. The calibration value and sign should 

match the 

ZERO:

 

nnnn 

value and 

ZERO: +/-: xxx

 value shown on the 

Calibration Detail 

label affixed to the pipe fitting. 

These are factory calibration factors that should not be changed without first 
contacting the factory for technical assistance.

6.3.2.8
Zero Off-Set
Setting

Zero Off-set

Sign

FR Offset

0.0000

Enter Value

Negative

Positive

The S4 will monitor the sound speed (SOS) of the measured fluid and report FAULT 12 if the 
measured SOS exceeds the 

SOS Fault Window 

percentage value. The speed that sound 

travels through the fluid will change if the density of the fluid changes (temperature change) 
or if the fluid composition changes (chemical concentration, different chemicals, etc.). The 
default value is 10%. The maximum allowable value is 50%.

6.3.2.8
SOS 
(Speed Of Sound)
Fault Window

SOS Fault Window

10

Enter Value

CALIBRATION DETAIL LABEL

Industries, Ltd.

Industries, Ltd.

ProSeries

by Blue-White Ind.

d

Industries, Ltd.

Industries, Ltd.

ProSeries

by Blue-White Ind.

d

Содержание ProSeries SONIC-PRO S4 Series

Страница 1: ...Installation and Operating Manual 2 TWO YEAR WARRANTY ProSeries Series S4 Ultrasonic Flow Meter Industries Ltd Industries Ltd ProSeries by Blue White Ind d SONIC PRO d...

Страница 2: ...DING THE SAFE USE OF THIS PRODUCT AND OTHER TECHNICAL ASSISTANCE MAY BE DIRECTED TO Blue White Industries 714 893 8529 techsupport blue white com CAUTION risk of electric shock CAUTION risk of danger...

Страница 3: ...ions of other chemicals or other chemicals entirely may be acceptable provided the chemical does not change the sound speed SOS more than approximately 100 meters per second and provided the fluid all...

Страница 4: ...must not be cut or modified Note that the Sonic Pro can accurately measure flow from either direction The SPU can be mounted on a wall or on a horizontal or vertical run of pipe Although the Sonic Pro...

Страница 5: ...is required on downward flows to ensure a full pipe See the minimum straight length of pipe requirement chart above Related piping must be independently supported The meter should not be allowed to s...

Страница 6: ...d tight cable gland connectors for cable diameter from 200 to 394 inches 5 1 to 10 0 mm Two liquid tight cable gland connectors cable for diameters from 118 to 255 inches 3 0 to 6 5 mm One communicati...

Страница 7: ...ector T2 1 2 Positive Negative Type Open collector w internal 1 K pull up to 5V Can be pulled up to 30V 10mA 3K pull up Duty Cycle 50 5 Frequency Range 0 1000 Hz Resolution 1 Hz Accuracy 0 5 F S PADDL...

Страница 8: ...allowable SOS Speed Of Sound error set point has been exceeded FAULT 21 The configuration data is not valid Flow Direction Total Units U S Gallons Min 105 1482 25 05 469 U S Gallons 0 28 Transducer S...

Страница 9: ...he value Press ENTER to confirm the modifications and return to the menu heading Press ESC to cancel the modifications and return to the menu heading 5 3 2 Selecting Optional Items Press up arrow p or...

Страница 10: ...use your inlet outlet pipe material 10 Fluid Type Select WATER or CUSTOM 11 Fluid SOS If WATER is selected it is not necessary to input a fluid SOS If CUSTOM fluid type is selected the Speed of Sound...

Страница 11: ...e factory set inside the pipe fitting and cannot be moved Select the mount method that is shown on the Calibration Detail label affixed to the pipe fitting Transducer Model I C J Mount Method W V Cabl...

Страница 12: ...ings Direction Zero A B Scaling A B Enter Value Sign A B Zero Value A B Positive B A A B Enter Value Zero B A Scaling B A Enter Value Sign B A Zero Value B A Enter Value 0 975 0 455 Negative Positive...

Страница 13: ...n the acceptable range and there are not too many particles or other sound blocking or reflecting contaminants in the fluid which can inhibit the sound beam The target Fluid SOS value can be adjusted...

Страница 14: ...els Oil Cubic Feet Acre Feet Imperial Gallons Cubic Meters Liters Milliliters FR Custom V Gal Fl Total Decimal 0 Enter Value Flow Average s 0 Enter Value Enable Tot Clear Yes No Low Fl Cut Off 0 Enter...

Страница 15: ...into any Tot Set Point 1 2 3 4 or 5 a new log will be generated when the total accumulated flow value is equal to the configured value To access the log entries you must copy the logs onto a USB flash...

Страница 16: ...ters Milliliters TOT Custom V Gal Fl Total Decimal 0 Enter Value Flow Average s 0 Enter Value Enable Tot Clear Yes No Low Fl Cut Off 0 Enter Value High Fl Cut Off 999 Enter Value Scaling Off set 1 000...

Страница 17: ...Off value totalizing will still occur 6 3 2 6 High Flow Rate Cut off Setting High Fl Cut Off 999 Enter Value Scaling Off Set is a multiplier used to correct for flow inaccuracies All flow rate values...

Страница 18: ...e meter was sleeping using the following formula Total flow display value T1 F1 F2 2 x N 1 Where T1 Last total before sleep F1 Last flow rate units per second before sleep F2 Flow rate average of the...

Страница 19: ...gh Release A 0 Enter Value Low Trigger A 0 Enter Value Low Release A 0 Enter Value Time s A 0 Enter Value Monitor Totalizer Batch Amt A 0 Enter Value Contact Timer s A 0 Enter Value Faults A Relay B M...

Страница 20: ...ue Rate at Max Curr 0 0000 Enter Value Max Analog O P 20 Enter Value 6 4 2 Pulse 0 1000 Hz Output Menu A digital pulse output signal can be programmed within the range of 0 1000 Hz pulses per second S...

Страница 21: ...ue Faults B None A None B Relay B Monitor Flow Rate High Trigger B 0 Enter Value High Release B 0 Enter Value Low Trigger B 0 Enter Value Low Release B 0 Enter Value Time s B 0 Enter Value None B 6 4...

Страница 22: ...Value Contact Timer s B 0 Enter Value 6 4 3 2 CC Assigned to Monitor FAULTS When in the Relay A or Relay B menu selecting and then exiting the FAULTS menu will assign that relay to energize when any...

Страница 23: ...unsigned char 8 Frequency output 45 Firmware Version unsigned char 8 8 ASCII characters A11 22 33 53 Log Entry unsigned char 86 Log Entry String Eg 2013 12 29 09 15 00 FR 0 0000 TL 0 00 Description P...

Страница 24: ...e meter can be re calibrated in the field by selecting Calibration and pressing the right arrow u or ENTER button The pipe must be full of fluid and there must be no movement of the fluid in the pipe...

Страница 25: ...Save Config function is possible at the main menu Global Config Password 1234 Enter Value Change Password Change Password Old Pass 1234 OK CANCEL NEW Pass 5678 OK Enter master password 3408 to clear a...

Страница 26: ...me Units Independently configurable Rate and Total display units in U S Gallons ounces barrels US liquid barrels US oil cubic ft acre ft Imperial British gallons liter cubic meter or user defined cust...

Страница 27: ...ipe fitting transducer no transducer cables Standard 4 20mA and High Speed Pulse output only Without power supply customer supplies 15 30 VDC 0 25 ft cable length with field replaceable liquid tight c...

Страница 28: ...03 5 41 1427 10 50 1447 20 68 1481 30 86 1507 40 104 1526 50 122 1541 60 140 1552 70 158 1555 80 176 1555 90 194 1550 100 212 1543 Water Sound Speeds 7 2 Sound speed data Page 27 Sonic Pro Industries...

Страница 29: ...in 18 39cm 5 90in 14 97cm 74in 1 88cm 2 75in 6 99cm 7 24in 18 39cm 5 83in 14 80cm 55in 1 39cm 2 75in 6 99cm Sonic Pro Inline Pipe Fittings Pipe Size 2 1 5 1 75 5 Pipe Size OL OW A B C 2 7 47in 18 97cm...

Страница 30: ...ers TOT Custom V Gal Fl Total Decimal 0 Enter Value Flow Average s 0 Enter Value Enable Tot Clear Yes No Low Fl Cut Off 0 Enter Value High Fl Cut Off 999 Enter Value Scaling Off set 1 0000 Enter Value...

Страница 31: ...Enter Value Time s B 0 Enter Value Monitor Totalizer Batch Amt B 0 Enter Value Contact Timer s B 0 Enter Value Faults B Clock CLOCK SET 2013 07 24 14 31 OK CANCEL Calibration Clear Cal Value None A N...

Страница 32: ...kmanship for 24 months from date of factory shipment WARRANTY COVERAGE IS LIMITED TO REPAIR OR REPLACEMENT OF THE DEFECTIVE FLOWMETER ONLY UNDER NO CIRCUMSTANCES SHALL BLUE WHITE BE LIABLE FOR ANY CON...

Отзывы: