Page 15
Sonic-Pro
6.3.2
Display
Menu
In the
DISPLAY
menu, the following features may be configured:
!
Flow Rate - Volume Units
!
Flow Rate - Time Units
!
Flow Rate - Decimal Point Location
!
Totalizer - Volume Units
!
Totalizer - Decimal Point Location
!
Flow Rate Averaging
!
Enable/Disable clearing the totalizer from the front panel
!
Low Flow Rate Cut-Off
!
High Flow Rate Cut-Off
!
Scaling Offset
!
Zero Offset
!
SOS Fault Window
!
Sleep Time
Display
Flow Rate
Volume Units
Ounces
U.S. Gallons
U.S. Barrels Liq.
U.S. Barrels Oil
Cubic Feet
Acre Feet
Imperial Gallons
Cubic Meters
Liters
Milliliters
FR Custom V/Gal
Time Units
Seconds
Hours
Days
Minutes
Fl. Rate Decimal
0
Enter Value
Totalizer
Volume Units
Ounces
U.S. Gallons
U.S. Barrels Liq.
U.S. Barrels Oil
Cubic Feet
Acre Feet
Imperial Gallons
Cubic Meters
Liters
Milliliters
TOT Custom V/Gal
Fl. Total Decimal
0
Enter Value
Flow Average(s)
0
Enter Value
Enable Tot. Clear?
Yes
No
Low Fl. Cut-Off
0
Enter Value
High Fl. Cut-Off
999
Enter Value
Scaling Off-set
1.0000
Enter Value
Zero Off-set
Sign
FR Offset
0.0000
Enter Value
Negative
Positive
SOS Fault Window
10
Enter Value
Sleep Time(s)
0
Enter Value
In the
Flow Rate
menu
,
the
Volume Units
and
Time Units
may be selected from a list.
A custom volume unit may be created by selecting
FR Custom V/Gal
and entering the
number of gallons in the custom volume.
The
Fl. Rate Decimal
decimal point location may be set to a maximum of 5 digits.
6.3.2.1
Flow Rate Menu
Flow Rate
Volume Units
Ounces
U.S. Gallons
U.S. Barrels Liq.
U.S. Barrels Oil
Cubic Feet
Acre Feet
Imperial Gallons
Cubic Meters
Liters
Milliliters
FR Custom V/Gal
Time Units
Seconds
Hours
Days
Minutes
Fl. Rate Decimal
0
Enter Value
0
Enter Value
0
Enter Value
0
Enter Value
Page 16
Sonic-Pro
In the
Totalizer
menu, the
Volume Units
may be selected from a list.
A custom volume unit may be created by selecting
FR Custom V/Gal
and entering the
number of gallons in the custom volume.
The
Fl. Total Decimal
decimal point location may be set to a maximum of 5 digits.
6.3.2.2
Totalizer Menu
Totalizer
Volume Units
Ounces
U.S. Gallons
U.S. Barrels Liq.
U.S. Barrels Oil
Cubic Feet
Acre Feet
Imperial Gallons
Cubic Meters
Liters
Milliliters
TOT Custom V/Gal
Fl. Total Decimal
0
Enter Value
0
Enter Value
Flow Average(s)
sets the running average period in seconds. Any number of seconds from 2
to 10 seconds is possible.
6.3.2.3
Flow Average
Setting
Flow Average(s)
0
Enter Value
Enable Tot. Clear?
enables or disables the ability of the user to clear the accumulated total
flow by pressing the ESC button.
6.3.2.4
Clear Total
Setting
Enable Tot. Clear?
Yes
No
Low Fl. Cut-Off
sets the minimum displayed flow rate. Calculated flow rate values less than
this value will be displayed as Zero and flow totalizing will not occur.
6.3.2.5
Low Flow Rate
Cut-off Setting
Low Fl. Cut-Off
0
Enter Value
High Fl. Cut-Off
sets the maximum displayed flow rate. Calculated flow rate values greater
than this value will be displayed as the
High Fl. Cut-Off value
, totalizing will still occur.
6.3.2.6
High Flow Rate
Cut-off Setting
High Fl. Cut-Off
999
Enter Value
Scaling Off-Set
is a multiplier used to correct for flow inaccuracies. All flow rate values will
be multiplied by the
Scaling Off-Set
value and the result displayed and totaled.
6.3.2.7
Scaling Off-Set
Setting
Scaling Off-set
1.0000
Enter Value
Zero Off-Set
is a factory calibration value used to offset the zero flow setpoint during the
initial calibration of the
Ultrasonic Inline Pipe Fitting
. The calibration value and sign should
match the
ZERO:
nnnn
value and
ZERO: +/-: xxx
value shown on the
Calibration Detail
label affixed to the pipe fitting.
These are factory calibration factors that should not be changed without first
contacting the factory for technical assistance.
6.3.2.8
Zero Off-Set
Setting
Zero Off-set
Sign
FR Offset
0.0000
Enter Value
Negative
Positive
The S4 will monitor the sound speed (SOS) of the measured fluid and report FAULT 12 if the
measured SOS exceeds the
SOS Fault Window
percentage value. The speed that sound
travels through the fluid will change if the density of the fluid changes (temperature change)
or if the fluid composition changes (chemical concentration, different chemicals, etc.). The
default value is 10%. The maximum allowable value is 50%.
6.3.2.8
SOS
(Speed Of Sound)
Fault Window
SOS Fault Window
10
Enter Value