Blue-White ProSeries SONIC-PRO S4 Series Скачать руководство пользователя страница 13

Page 11

Sonic-Pro

The relative flow direction is shown on the 

Calibration Detail

 label affixed to the pipe fitting.

6.2.2.4
Fluid Flow
Direction

Each of the flow directions (A to B and B to A) have a 

Scaling Factor 

and either a positive or 

negative 

Zero Factor 

associated with it. The correct factors are shown on the

 Calibration 

Detail

 label affixed to the ultrasonic flow sensor pipe fitting. 

These are factory calibration factors that should not be changed without first 
contacting the factory for technical assistance.

Transducer

Model

Mount Method

Cable Length(ft)

Flow Direction

Direction

B -> A

A -> B

Transducer

Model

Mount Method

Cable Length(ft)

Flow Direction

Settings

Direction

Zero(A->B)

Scaling(A->B)

Enter Value

Sign(A->B)

Zero Value(A->B)

Positive

B -> A

A -> B

Enter Value

Zero(B->A)

Scaling(B->A)

Enter Value

Sign(B->A)

Zero Value(B->A)

Enter Value

0.975

0.455

Negative

Positive

Negative

0.975

0.455

6.2.2.5
Fluid Flow
Direction

(CAL Settings)

6.2.3
Pipe
Menu

Pipe

Pipe Outside Dia.

Enter Value

Pipe Wall Thick.

Enter Value

Pipe Material

Custom

Pipe SOS

0.0000

Enter Value

0.0000

0.0000

HDPE

Each pipe fitting has a specific 

Pipe Outside Dia., Pipe Wall Thick., Pipe Material 

and 

Pipe 

SOS 

(Speed Of Sound) associated with it. The correct factors are shown on the

 Calibration 

Detail

 label affixed to the ultrasonic flow sensor pipe fitting. 

These are factory calibration factors that should not be changed without first 
contacting the factory for technical assistance.

CALIBRATION DETAIL LABEL

Page 12

Sonic-Pro

 

6.2.4
Fluid
Menu

Fluid

Fluid Type

Fluid SOS

Water

Custom

0000

Enter Value

The 

sound transducers are located in the S4 pipe fitting. They are carefully positioned at the 

factory for use with water at 68 degrees F which has a SOS (Speed Of Sound) of 1481 
meters per second. (see page 28 for water SOS chart). The transducer placement cannot be 
adjusted in the field. As the fluid’s sound speed moves away from the target 

Fluid SOS

measurements become more difficult. If the speed that sound travels through the fluid is very 
different than that of water, the meter will not function. The acceptable SOS range is from 
1,400 to 1,600 m/sec.

The meter will display the SOS of the fluid being measured in the upper right hand corner of 
the display. 

Successful measurements are possible with fluids other than water provided the fluid’s SOS 
is within the acceptable range and there are not too many particles or other sound blocking or 
reflecting contaminants in the fluid which can inhibit the sound beam.

The target 

Fluid SOS 

value can be adjusted to maximize the meter’s ability to measure the 

fluid. The chart below provides the sound speeds of a number of chemicals of specific 
concentrations at specific temperatures. Diluted chemicals will have a different sound speed. 
Select 

Custom 

from the 

Fluid Type 

menu, then input the SOS fluid in the 

Fluid SOS 

menu. 

Fluid

Water

o

o

Temp.  F ( C)

32 (0)

Sound Speed (meters/sec)

1403

Fluid Sound Speeds

Water

68 (20)

1481

Water

212 (100)

1543

1520

Sea Water

68 (20)

1900

Aluminum Sulfate 10% (Alum)

68 (20)

1578

Ethylene Glycol (50% in water)

77 (25)

1250

Gasoline

77 (25)

1581

Hydrochloric Acid 32%

1530

Hydrogen Peroxide 31%

1548

Milk

77 (25)

1250

Oil, Diesel

77 (25)

1487

Oil, Motor (SAE 30)

1431

Oil, Mineral

77 (25)

2800

Potassium Hydroxide 50%

68 (20)

68 (20)

68 (20)

68 (20)

1835

Sodium Hypochlorite 12%

68 (20)

1260

Sulfuric Acid 96%

68 (20)

Industries, Ltd.

Industries, Ltd.

ProSeries

by Blue-White Ind.

d

Industries, Ltd.

Industries, Ltd.

ProSeries

by Blue-White Ind.

d

Содержание ProSeries SONIC-PRO S4 Series

Страница 1: ...Installation and Operating Manual 2 TWO YEAR WARRANTY ProSeries Series S4 Ultrasonic Flow Meter Industries Ltd Industries Ltd ProSeries by Blue White Ind d SONIC PRO d...

Страница 2: ...DING THE SAFE USE OF THIS PRODUCT AND OTHER TECHNICAL ASSISTANCE MAY BE DIRECTED TO Blue White Industries 714 893 8529 techsupport blue white com CAUTION risk of electric shock CAUTION risk of danger...

Страница 3: ...ions of other chemicals or other chemicals entirely may be acceptable provided the chemical does not change the sound speed SOS more than approximately 100 meters per second and provided the fluid all...

Страница 4: ...must not be cut or modified Note that the Sonic Pro can accurately measure flow from either direction The SPU can be mounted on a wall or on a horizontal or vertical run of pipe Although the Sonic Pro...

Страница 5: ...is required on downward flows to ensure a full pipe See the minimum straight length of pipe requirement chart above Related piping must be independently supported The meter should not be allowed to s...

Страница 6: ...d tight cable gland connectors for cable diameter from 200 to 394 inches 5 1 to 10 0 mm Two liquid tight cable gland connectors cable for diameters from 118 to 255 inches 3 0 to 6 5 mm One communicati...

Страница 7: ...ector T2 1 2 Positive Negative Type Open collector w internal 1 K pull up to 5V Can be pulled up to 30V 10mA 3K pull up Duty Cycle 50 5 Frequency Range 0 1000 Hz Resolution 1 Hz Accuracy 0 5 F S PADDL...

Страница 8: ...allowable SOS Speed Of Sound error set point has been exceeded FAULT 21 The configuration data is not valid Flow Direction Total Units U S Gallons Min 105 1482 25 05 469 U S Gallons 0 28 Transducer S...

Страница 9: ...he value Press ENTER to confirm the modifications and return to the menu heading Press ESC to cancel the modifications and return to the menu heading 5 3 2 Selecting Optional Items Press up arrow p or...

Страница 10: ...use your inlet outlet pipe material 10 Fluid Type Select WATER or CUSTOM 11 Fluid SOS If WATER is selected it is not necessary to input a fluid SOS If CUSTOM fluid type is selected the Speed of Sound...

Страница 11: ...e factory set inside the pipe fitting and cannot be moved Select the mount method that is shown on the Calibration Detail label affixed to the pipe fitting Transducer Model I C J Mount Method W V Cabl...

Страница 12: ...ings Direction Zero A B Scaling A B Enter Value Sign A B Zero Value A B Positive B A A B Enter Value Zero B A Scaling B A Enter Value Sign B A Zero Value B A Enter Value 0 975 0 455 Negative Positive...

Страница 13: ...n the acceptable range and there are not too many particles or other sound blocking or reflecting contaminants in the fluid which can inhibit the sound beam The target Fluid SOS value can be adjusted...

Страница 14: ...els Oil Cubic Feet Acre Feet Imperial Gallons Cubic Meters Liters Milliliters FR Custom V Gal Fl Total Decimal 0 Enter Value Flow Average s 0 Enter Value Enable Tot Clear Yes No Low Fl Cut Off 0 Enter...

Страница 15: ...into any Tot Set Point 1 2 3 4 or 5 a new log will be generated when the total accumulated flow value is equal to the configured value To access the log entries you must copy the logs onto a USB flash...

Страница 16: ...ters Milliliters TOT Custom V Gal Fl Total Decimal 0 Enter Value Flow Average s 0 Enter Value Enable Tot Clear Yes No Low Fl Cut Off 0 Enter Value High Fl Cut Off 999 Enter Value Scaling Off set 1 000...

Страница 17: ...Off value totalizing will still occur 6 3 2 6 High Flow Rate Cut off Setting High Fl Cut Off 999 Enter Value Scaling Off Set is a multiplier used to correct for flow inaccuracies All flow rate values...

Страница 18: ...e meter was sleeping using the following formula Total flow display value T1 F1 F2 2 x N 1 Where T1 Last total before sleep F1 Last flow rate units per second before sleep F2 Flow rate average of the...

Страница 19: ...gh Release A 0 Enter Value Low Trigger A 0 Enter Value Low Release A 0 Enter Value Time s A 0 Enter Value Monitor Totalizer Batch Amt A 0 Enter Value Contact Timer s A 0 Enter Value Faults A Relay B M...

Страница 20: ...ue Rate at Max Curr 0 0000 Enter Value Max Analog O P 20 Enter Value 6 4 2 Pulse 0 1000 Hz Output Menu A digital pulse output signal can be programmed within the range of 0 1000 Hz pulses per second S...

Страница 21: ...ue Faults B None A None B Relay B Monitor Flow Rate High Trigger B 0 Enter Value High Release B 0 Enter Value Low Trigger B 0 Enter Value Low Release B 0 Enter Value Time s B 0 Enter Value None B 6 4...

Страница 22: ...Value Contact Timer s B 0 Enter Value 6 4 3 2 CC Assigned to Monitor FAULTS When in the Relay A or Relay B menu selecting and then exiting the FAULTS menu will assign that relay to energize when any...

Страница 23: ...unsigned char 8 Frequency output 45 Firmware Version unsigned char 8 8 ASCII characters A11 22 33 53 Log Entry unsigned char 86 Log Entry String Eg 2013 12 29 09 15 00 FR 0 0000 TL 0 00 Description P...

Страница 24: ...e meter can be re calibrated in the field by selecting Calibration and pressing the right arrow u or ENTER button The pipe must be full of fluid and there must be no movement of the fluid in the pipe...

Страница 25: ...Save Config function is possible at the main menu Global Config Password 1234 Enter Value Change Password Change Password Old Pass 1234 OK CANCEL NEW Pass 5678 OK Enter master password 3408 to clear a...

Страница 26: ...me Units Independently configurable Rate and Total display units in U S Gallons ounces barrels US liquid barrels US oil cubic ft acre ft Imperial British gallons liter cubic meter or user defined cust...

Страница 27: ...ipe fitting transducer no transducer cables Standard 4 20mA and High Speed Pulse output only Without power supply customer supplies 15 30 VDC 0 25 ft cable length with field replaceable liquid tight c...

Страница 28: ...03 5 41 1427 10 50 1447 20 68 1481 30 86 1507 40 104 1526 50 122 1541 60 140 1552 70 158 1555 80 176 1555 90 194 1550 100 212 1543 Water Sound Speeds 7 2 Sound speed data Page 27 Sonic Pro Industries...

Страница 29: ...in 18 39cm 5 90in 14 97cm 74in 1 88cm 2 75in 6 99cm 7 24in 18 39cm 5 83in 14 80cm 55in 1 39cm 2 75in 6 99cm Sonic Pro Inline Pipe Fittings Pipe Size 2 1 5 1 75 5 Pipe Size OL OW A B C 2 7 47in 18 97cm...

Страница 30: ...ers TOT Custom V Gal Fl Total Decimal 0 Enter Value Flow Average s 0 Enter Value Enable Tot Clear Yes No Low Fl Cut Off 0 Enter Value High Fl Cut Off 999 Enter Value Scaling Off set 1 0000 Enter Value...

Страница 31: ...Enter Value Time s B 0 Enter Value Monitor Totalizer Batch Amt B 0 Enter Value Contact Timer s B 0 Enter Value Faults B Clock CLOCK SET 2013 07 24 14 31 OK CANCEL Calibration Clear Cal Value None A N...

Страница 32: ...kmanship for 24 months from date of factory shipment WARRANTY COVERAGE IS LIMITED TO REPAIR OR REPLACEMENT OF THE DEFECTIVE FLOWMETER ONLY UNDER NO CIRCUMSTANCES SHALL BLUE WHITE BE LIABLE FOR ANY CON...

Отзывы: