Sonic-Pro

Page 24

Sonic-Pro

6.5
Configure the
Clock

In the 

Clock menu, 

use the right arrow 

u

 or left arrow 

t

 to highlight the Year, Month, Day, 

hour, or minute. Change the highlighted value using the up arrow 

p

 or down arrow 

q

 button. 

To sve the entries, highlight 

OK

 and press 

ENTER.

The clock is powered by a supercapacitor, which is continuously charging while the meter is 
powered, rather than a battery. It never has to be replaced.

Clock

CLOCK SET

2013/07/24

14:31

OK

CANCEL

6.6
Zero Flow
Calibration

The zero calibration data can be stored for one flow direction only. If the 

Pipe Fitting 

and 

SPU 

was shipped from the factory as a single model number, the zero flow 

Calibration 

required to link the SPU and the Pipe Fitting was performed at the factory for the A --> B flow 
direction. No further calibration is necessary. If the flow direction is from A <--- B or if a new 

Ultrasonic Inline Pipe Fitting 

is used, re-calibration may be necessary to achieve high 

accuracy.

The meter can be re-calibrated in the field by selecting 

Calibration

 and pressing the right 

arrow 

u

 or 

ENTER

 button. The pipe must be full of fluid and there must be no movement of 

the fluid in the pipe during the calibration process. 

Note that performing a 

Zero Flow

 

Calibration 

does not affect the Flow Direction 

configuration settings shown on the Calibration Detail Label:  

Zero 

and 

Zero +/- 

 variables: 

transducer/Flow Direction/Settings/A -> B/ 

Calibration

Clear Cal. Value

CALIBRATION

CANCEL

Yes

No

Calibration

6.7
Global
Configuration

In the 

Global Config. 

menu, the user can configure a password, restore the factory default 

configuration data, save and restore configuration data to a USB thumb drive, and update or 
view the current firmware version.

6.7.1
Password

The S4 is shipped from the factory with the default password 0000. If a new password is not 
configured, the meter will allow saving changes to the configuration at the main menu. 

Configuration

Run

Save Config

A user defined four digit password can be entered to replace 0000. 

Change Password

Change Password

New Pass.:

1234

CANCEL

Confirm Pass.: 1234

Once entered, the user defined password can be changed to any four digit number.

Once a new password has been configured, it must be entered to allow saving any changes 
made to the configuration data. Once entered, the 

Save Config 

function is possible at the 

main menu. 

Global Config.

Password

1234

Enter Value

Change Password

Change Password

Old Pass.:

1234

OK

CANCEL

NEW Pass.:

5678

OK

Enter master password 3408 to clear any stored password.

6.7.2
Save/Restore
Config

The S4 configuration settings can be save and restored using a USB flash drive.

Save Conf. To USB

USB Drive Found!
Config. Copied To USB

Save Conf. To USB

OK

Restore USB Conf.

Config. File Found.
Erasing Config File..

Restore USB Conf.

Restored Config.

OK

6.7.3
Firmware
Update

The S4 firmware can be upgraded to the latest version. Download from 

 

www.blue-white.com

to a USB flash drive. 

Configuration data will not be erased. 

Global Config.

Password

0

Enter Value

Firmware Update

Change Password

Change Password

Old Pass.:

0000

OK

CANCEL

NEW Pass.:

0000

Save Conf. To USB

USB Drive Found!
Config. Copied To USB

Save Conf. To USB

OK

Restore USB Conf.

Config. File Found.
Erasing Config File..

Restore USB Conf.

Restored Config.

OK

USB Drive Found.
Press ESC to upgrade

OK

Firmware Upgrade

Firmware Update

USB Drive Found.
Press ESC to upgrade

OK

Firmware Upgrade

Industries, Ltd.

Industries, Ltd.

ProSeries

by Blue-White Ind.

d

Industries, Ltd.

Industries, Ltd.

ProSeries

by Blue-White Ind.

d

Содержание ProSeries SONIC-PRO S4 Series

Страница 1: ...Installation and Operating Manual 2 TWO YEAR WARRANTY ProSeries Series S4 Ultrasonic Flow Meter Industries Ltd Industries Ltd ProSeries by Blue White Ind d SONIC PRO d...

Страница 2: ...DING THE SAFE USE OF THIS PRODUCT AND OTHER TECHNICAL ASSISTANCE MAY BE DIRECTED TO Blue White Industries 714 893 8529 techsupport blue white com CAUTION risk of electric shock CAUTION risk of danger...

Страница 3: ...ions of other chemicals or other chemicals entirely may be acceptable provided the chemical does not change the sound speed SOS more than approximately 100 meters per second and provided the fluid all...

Страница 4: ...must not be cut or modified Note that the Sonic Pro can accurately measure flow from either direction The SPU can be mounted on a wall or on a horizontal or vertical run of pipe Although the Sonic Pro...

Страница 5: ...is required on downward flows to ensure a full pipe See the minimum straight length of pipe requirement chart above Related piping must be independently supported The meter should not be allowed to s...

Страница 6: ...d tight cable gland connectors for cable diameter from 200 to 394 inches 5 1 to 10 0 mm Two liquid tight cable gland connectors cable for diameters from 118 to 255 inches 3 0 to 6 5 mm One communicati...

Страница 7: ...ector T2 1 2 Positive Negative Type Open collector w internal 1 K pull up to 5V Can be pulled up to 30V 10mA 3K pull up Duty Cycle 50 5 Frequency Range 0 1000 Hz Resolution 1 Hz Accuracy 0 5 F S PADDL...

Страница 8: ...allowable SOS Speed Of Sound error set point has been exceeded FAULT 21 The configuration data is not valid Flow Direction Total Units U S Gallons Min 105 1482 25 05 469 U S Gallons 0 28 Transducer S...

Страница 9: ...he value Press ENTER to confirm the modifications and return to the menu heading Press ESC to cancel the modifications and return to the menu heading 5 3 2 Selecting Optional Items Press up arrow p or...

Страница 10: ...use your inlet outlet pipe material 10 Fluid Type Select WATER or CUSTOM 11 Fluid SOS If WATER is selected it is not necessary to input a fluid SOS If CUSTOM fluid type is selected the Speed of Sound...

Страница 11: ...e factory set inside the pipe fitting and cannot be moved Select the mount method that is shown on the Calibration Detail label affixed to the pipe fitting Transducer Model I C J Mount Method W V Cabl...

Страница 12: ...ings Direction Zero A B Scaling A B Enter Value Sign A B Zero Value A B Positive B A A B Enter Value Zero B A Scaling B A Enter Value Sign B A Zero Value B A Enter Value 0 975 0 455 Negative Positive...

Страница 13: ...n the acceptable range and there are not too many particles or other sound blocking or reflecting contaminants in the fluid which can inhibit the sound beam The target Fluid SOS value can be adjusted...

Страница 14: ...els Oil Cubic Feet Acre Feet Imperial Gallons Cubic Meters Liters Milliliters FR Custom V Gal Fl Total Decimal 0 Enter Value Flow Average s 0 Enter Value Enable Tot Clear Yes No Low Fl Cut Off 0 Enter...

Страница 15: ...into any Tot Set Point 1 2 3 4 or 5 a new log will be generated when the total accumulated flow value is equal to the configured value To access the log entries you must copy the logs onto a USB flash...

Страница 16: ...ters Milliliters TOT Custom V Gal Fl Total Decimal 0 Enter Value Flow Average s 0 Enter Value Enable Tot Clear Yes No Low Fl Cut Off 0 Enter Value High Fl Cut Off 999 Enter Value Scaling Off set 1 000...

Страница 17: ...Off value totalizing will still occur 6 3 2 6 High Flow Rate Cut off Setting High Fl Cut Off 999 Enter Value Scaling Off Set is a multiplier used to correct for flow inaccuracies All flow rate values...

Страница 18: ...e meter was sleeping using the following formula Total flow display value T1 F1 F2 2 x N 1 Where T1 Last total before sleep F1 Last flow rate units per second before sleep F2 Flow rate average of the...

Страница 19: ...gh Release A 0 Enter Value Low Trigger A 0 Enter Value Low Release A 0 Enter Value Time s A 0 Enter Value Monitor Totalizer Batch Amt A 0 Enter Value Contact Timer s A 0 Enter Value Faults A Relay B M...

Страница 20: ...ue Rate at Max Curr 0 0000 Enter Value Max Analog O P 20 Enter Value 6 4 2 Pulse 0 1000 Hz Output Menu A digital pulse output signal can be programmed within the range of 0 1000 Hz pulses per second S...

Страница 21: ...ue Faults B None A None B Relay B Monitor Flow Rate High Trigger B 0 Enter Value High Release B 0 Enter Value Low Trigger B 0 Enter Value Low Release B 0 Enter Value Time s B 0 Enter Value None B 6 4...

Страница 22: ...Value Contact Timer s B 0 Enter Value 6 4 3 2 CC Assigned to Monitor FAULTS When in the Relay A or Relay B menu selecting and then exiting the FAULTS menu will assign that relay to energize when any...

Страница 23: ...unsigned char 8 Frequency output 45 Firmware Version unsigned char 8 8 ASCII characters A11 22 33 53 Log Entry unsigned char 86 Log Entry String Eg 2013 12 29 09 15 00 FR 0 0000 TL 0 00 Description P...

Страница 24: ...e meter can be re calibrated in the field by selecting Calibration and pressing the right arrow u or ENTER button The pipe must be full of fluid and there must be no movement of the fluid in the pipe...

Страница 25: ...Save Config function is possible at the main menu Global Config Password 1234 Enter Value Change Password Change Password Old Pass 1234 OK CANCEL NEW Pass 5678 OK Enter master password 3408 to clear a...

Страница 26: ...me Units Independently configurable Rate and Total display units in U S Gallons ounces barrels US liquid barrels US oil cubic ft acre ft Imperial British gallons liter cubic meter or user defined cust...

Страница 27: ...ipe fitting transducer no transducer cables Standard 4 20mA and High Speed Pulse output only Without power supply customer supplies 15 30 VDC 0 25 ft cable length with field replaceable liquid tight c...

Страница 28: ...03 5 41 1427 10 50 1447 20 68 1481 30 86 1507 40 104 1526 50 122 1541 60 140 1552 70 158 1555 80 176 1555 90 194 1550 100 212 1543 Water Sound Speeds 7 2 Sound speed data Page 27 Sonic Pro Industries...

Страница 29: ...in 18 39cm 5 90in 14 97cm 74in 1 88cm 2 75in 6 99cm 7 24in 18 39cm 5 83in 14 80cm 55in 1 39cm 2 75in 6 99cm Sonic Pro Inline Pipe Fittings Pipe Size 2 1 5 1 75 5 Pipe Size OL OW A B C 2 7 47in 18 97cm...

Страница 30: ...ers TOT Custom V Gal Fl Total Decimal 0 Enter Value Flow Average s 0 Enter Value Enable Tot Clear Yes No Low Fl Cut Off 0 Enter Value High Fl Cut Off 999 Enter Value Scaling Off set 1 0000 Enter Value...

Страница 31: ...Enter Value Time s B 0 Enter Value Monitor Totalizer Batch Amt B 0 Enter Value Contact Timer s B 0 Enter Value Faults B Clock CLOCK SET 2013 07 24 14 31 OK CANCEL Calibration Clear Cal Value None A N...

Страница 32: ...kmanship for 24 months from date of factory shipment WARRANTY COVERAGE IS LIMITED TO REPAIR OR REPLACEMENT OF THE DEFECTIVE FLOWMETER ONLY UNDER NO CIRCUMSTANCES SHALL BLUE WHITE BE LIABLE FOR ANY CON...

Отзывы: