G5 Weighing Instrument
1-1
1. Introduction
General
The G5 Instruments are high performance single-channel weight indicators (PM model,
panel mounted) or weight transmitters (RM model, DIN rail mounted) intended for
industrial systems. This manual describes the functionality of the optional Fieldbus
interface of the G5 Instrument. The instrument supports the following busses:
ProfibusDP, DeviceNet, ControlNet and EtherNet/IP. The optional fieldbus module is
mounted in the Fieldbus slot.
l fieldbus settings are done with setup parameters in the instrument. No settings are
done on the module itself.
Module installation
Make sure the instrument power supply is off or disconnected. Remove the plastic
cover from the fieldbus slot. Insert the adaptor
very carefully and make absolutely
sure that the adaptor slides correctly into the guides
in the connector on the PCB.
Tighten the two fastening screws at the adaptor front and check that the two securing
hooks locks into the PCB.
Module type
Connector
Baud rate
Address
Bus Supply
Voltage
Profibus-DP
Profibus
9-pin, female
D-sub (DB9F)
9.6 kbps –
12 Mbps or
Auto set by
parameter
1 – 125,
set by parameter
DeviceNet
5 pin male
connector.
125, 250,
500 kbps or
Auto set by
parameter.
0 – 63,
set by parameter
According to
DeviceNet (Node)
specification:
nominal 24 VDC,
range 11 – 25
VDC.
ControlNet
BNC.
Fixed 5
Mbit/s.
1 – 99, set by
parameter
EtherNet/IP
Dual RJ45
(Only port 1
,P1) is
supported
Selectable Static
or DHCP. If static
choosen IP-
address, Net
Mask and Default
Gateway are set
by parameters.