G5 Weighing Instrument
6-3
Profibus DP-V1
Use the provided GDS file VISH0F83 with revision 2. This revision is intended for G5
with SW revision 1.3.0 or later. Revision 1 of the GDS file is used with G5 with SW
revision 1.0.0, 1.2.0 or 2.0.1 but cannot be used with 1.3.0 that is described in this
manual. This is because the mapping and amount of data exchanged is different.
After installing the GDS file select the modules as:
Output module 4 words
4 words = 8 bytes.
Input Module 4 words
Input Module 10 words
Input Module 16 words
Input Module 10 words
Input Module 5 words
Input Module 8 words
53 words = 106 bytes.
Each module has consistent data within the module. The modules sizes are chosen to
fit the different group in the mapping. Configuration must match exactly to be valid.
Note that the Profibus module will by default get the address 126 which is not valid but
will prevent the module from causing address collision when connected to the network.
Select the desired address.
DeviceNet
Use the 005A000000620200.eds file for the DeviceNet module if needed. Data sizes
should be selected for 8 bytes to the G5 instrument and 106 bytes from the instrument.
ControlNet
There are 2 eds files available for ControlNet,
EDS_ABCC_CNT_V_1_02.eds
and
EDS_PLC-5_ABCC_CNT_V_1_02.eds
. Make sure data sizes are set-up correctly.