Fieldbus Manual
3-2
Range/Alternatives
Explanation and
<default value>
result of alternatives.
DeviceNet, Address
Range 0 - 63
< 63 >
DeviceNet address setting
Note: This parameter is only shown if parameter
Fieldbus
(in
Hardware Configuration menu) is set to
DeviceNet
.
ControlNet, Address
Range 1 - 99
< 99 >
ControlNet address setting
Note: This parameter is only shown if parameter
Fieldbus
(in
Hardware Configuration menu) is set to
ControlNet
.
DeviceNet, Baudrate
125 kbps
250 kbps
500 kbps
Auto
< Auto >
The baud rate for a DeviceNet type of module. Must be selected
to suite the master.
125 kbps: Fixed baudrate 125 kbits/s.
250 kbps: Fixed baudrate 250 kbits/s.
500 kbps: Fixed baudrate 500 kbits/s.
Auto: Auto setting 125 – 500 kbits/s.
Note: This parameter is only shown if parameter
Fieldbus
(in
Hardware Configuration menu) is set to
DeviceNet
.
IP Address
< 192.168.99.151 >
The ip address of the EtherNet/IP interface.
Note: This parameter is only shown if parameter
Fieldbus
(in
Hardware Configuration menu) is set to
EtherNet/IP
. Restart the
instrument when the ip-address has been changed.
Net Mask
< 255.255.255.0 >
The net mask of the EtherNet/IP interface.
Note: This parameter is only shown if parameter
Fieldbus
(in
Hardware Configuration menu) is set to
EtherNet/IP
. Restart the
instrument when the net-mask has been changed.
Default Gateway
< 192.168.99.1 >
Default gateway of the EtherNet/IP interface.
Note: This parameter is only shown if parameter
Fieldbus
(in
Hardware Configuration menu) is set to
EtherNet/IP
. Restart the
instrument when the default gateway has been changed.
Data Format
Floating Point
Integer
< Floating Point >
Sets what data format should be used. The data format choice
applies to weight and some other values. Status, error code and
other similar data which is inherently integer data is always
given as integers (no decimals)
Floating Point: IEEE 32 bit floating point format.
Integer: 32 bit integer format.
Note: This parameter is not shown if parameter
Fieldbus
(in
Hardware Configuration menu) is set to
None
.