2.9 CHECKLIST FOR ANAESTHETIC EQUIPMENT 2003
The following checks should be made prior to each operating session. In addition,
checks 2, 6 and 9 (Monitoring, Breathing System and Ancillary Equipment) should
be made prior to each new patient during a session.
1. Check that the anaesthetic machine is connected to the electricity supply
(if appropriate) and switched on.
Note: Some anaesthetic workstations may enter an integral self-test programme when
switched on; those functions tested by such a programme need not be retested.
Take note of any information or labelling on the anaesthetic machine referring to the
current status of the machine. Particular attention should be paid to recent servicing.
Servicing labels should be fixed in the service logbook.
2. Check that all monitoring devices, in particular analyzer, pulse oximeter
and capnograph, are functioning and have appropriate alarm limits.
•
Check that gas sampling lines are properly attached and free of obstructions.
•
Check that an appropriate frequency of recording non-invasive blood pressure is
selected.
(Some monitors need to be in stand-by mode to avoid unnecessary alarms before being
connected to the patient).
Oxygen Monitor Test
The Oxygen Calibration menu allows adjustment of the displayed oxygen concentration
to match the gas that the probe is exposed to; this is normally 100% pure oxygen.
NOTE: Calibration is only possible if the ventilator is in Standby and a probe is
connected.
•
Press the Main menu key, select Oxygen Calibration.
•
Expose the probe to 100% pure oxygen flow or 21% room air.
•
Adjust the displayed value to 100% or 21% as applicable by turning and pressing the
Trak wheel.
•
Press the Trak wheel again to start the calibration process. This takes approximately one
minute and a countdown is displayed whilst it is in progress.
2. User Pre Use Testing
Содержание Sirius 1000
Страница 17: ...16 First Line and Planned Maintenance Manual Frontline Sirius Notes ...
Страница 27: ...26 First Line and Planned Maintenance Manual Frontline Sirius Notes 1 Technical Description ...
Страница 32: ...31 First Line and Planned Maintenance Manual Frontline Sirius 2 0 User Pre Use Tests 2 User Pre Use Testing ...
Страница 37: ...36 First Line and Planned Maintenance Manual Frontline Sirius Notes 2 User Pre Use Testing ...
Страница 50: ...49 First Line and Planned Maintenance Manual Frontline Sirius 3 0 Full System Test 3 Full System Testing ...
Страница 61: ...60 First Line and Planned Maintenance Manual Frontline Sirius Notes 3 Full System Testing ...
Страница 62: ...61 First Line and Planned Maintenance Manual Frontline Sirius 4 0 Maintenance 4 Maintenance ...
Страница 68: ...67 First Line and Planned Maintenance Manual Frontline Sirius 5 0 Routine Maintenance 5 Routine Maintenance ...
Страница 71: ...70 First Line and Planned Maintenance Manual Frontline Sirius Notes 5 Routine Maintenance ...
Страница 72: ...71 First Line and Planned Maintenance Manual Frontline Sirius 6 0 Planned Maintenance 6 Planned Maintenance ...
Страница 87: ...86 First Line and Planned Maintenance Manual Frontline Sirius Notes 7 Fitting Planned Maintenance Kits ...
Страница 95: ...94 First Line and Planned Maintenance Manual Frontline Sirius Regulators 7 Fitting Planned Maintenance Kits ...
Страница 105: ...104 First Line and Planned Maintenance Manual Frontline Sirius Notes 7 Fitting Planned Maintenance Kits ...
Страница 106: ...105 First Line and Planned Maintenance Manual Frontline Sirius 8 0 Spare Parts 8 Spare Parts ...