First Line and Planned Maintenance Manual
Frontline Sirius®
Warnings
The following statements are made to comply with the requirements of IEC 60601-1.
1. This equipment must only be connected to gas pipeline supply lines that are fitted with
pressure relief valves that limit the supply pressure to less than 7 bar.
2. The functioning of this machine may be adversely affected by the operation of equipment
such as high frequency surgical (diathermy) equipment, defibrillators or shortwave therapy
equipment in the vicinity. Increasing the distance from such equipment will minimise any
possible interference.
3. Prior to connecting the machine to a patient carry out the pre-use check to verify correct
alarm operation. To verify the O2 alarm, set the flowmeters to give a concentration of 50%
oxygen. Using the controls on the oxygen monitor panel, set the low oxygen level to 60%
and verify the oxygen low alarm operates. Set the high oxygen alarm level to 40% and
verify that the oxygen high alarm operates.
4. The oxygen flow can only be reduced to zero by turning the ON/OFF switch to the OFF
position. Excessive force on the oxygen control knob may damage the hypoxic guard.
5. To avoid explosion hazards, flammable anaesthetic agents such as ether and cyclopropane
must not be used in these machines. Only anaesthetic agents which comply with the
requirements on non-flammable anaesthetic agents in IEC 60601-2-13 ‘Specification for
Anaesthetic Machines’, are suitable for use in these machines.
6. As these machines are not suitable for use with flammable anaesthetic agents such as ether
and cyclopropane the use of antistatic breathing tubes and face masks is not necessary.
The use of antistatic or electrically conductive breathing tubes when utilising high frequency
surgery equipment may cause burns and is therefore not recommended in any application
that involves such apparatus.
7. The equipment must be periodically checked and maintained to ensure proper operation.
8. Performance of the equipment may be affected at temperatures below 10
o
C (50
o
F) and
above 40
o
C (104
o
F).
9. The performance of the anaesthetic machines and vaporizers may be degraded if the two
are mismatched. Refer to the vaporizer manufacturer’s instruction manual before use.
10. If the integrated oxygen analyser is not fitted, an oxygen analyser complying with ISO 7767
shall be used when the anaesthetic machine is in use.
11. The units use semiconductor devices which are susceptible to damage by overloading,
reversed polarity, electrostatic discharge and excessive heat or radiation. Avoid hazards
such as reversal of batteries, prolonged soldering, strong RF fields or other forms of
radiation, use of insulation testers or accidentally applied short circuits. Even the leakage
current from an unearthed soldering iron may cause trouble.
Содержание Sirius 1000
Страница 17: ...16 First Line and Planned Maintenance Manual Frontline Sirius Notes ...
Страница 27: ...26 First Line and Planned Maintenance Manual Frontline Sirius Notes 1 Technical Description ...
Страница 32: ...31 First Line and Planned Maintenance Manual Frontline Sirius 2 0 User Pre Use Tests 2 User Pre Use Testing ...
Страница 37: ...36 First Line and Planned Maintenance Manual Frontline Sirius Notes 2 User Pre Use Testing ...
Страница 50: ...49 First Line and Planned Maintenance Manual Frontline Sirius 3 0 Full System Test 3 Full System Testing ...
Страница 61: ...60 First Line and Planned Maintenance Manual Frontline Sirius Notes 3 Full System Testing ...
Страница 62: ...61 First Line and Planned Maintenance Manual Frontline Sirius 4 0 Maintenance 4 Maintenance ...
Страница 68: ...67 First Line and Planned Maintenance Manual Frontline Sirius 5 0 Routine Maintenance 5 Routine Maintenance ...
Страница 71: ...70 First Line and Planned Maintenance Manual Frontline Sirius Notes 5 Routine Maintenance ...
Страница 72: ...71 First Line and Planned Maintenance Manual Frontline Sirius 6 0 Planned Maintenance 6 Planned Maintenance ...
Страница 87: ...86 First Line and Planned Maintenance Manual Frontline Sirius Notes 7 Fitting Planned Maintenance Kits ...
Страница 95: ...94 First Line and Planned Maintenance Manual Frontline Sirius Regulators 7 Fitting Planned Maintenance Kits ...
Страница 105: ...104 First Line and Planned Maintenance Manual Frontline Sirius Notes 7 Fitting Planned Maintenance Kits ...
Страница 106: ...105 First Line and Planned Maintenance Manual Frontline Sirius 8 0 Spare Parts 8 Spare Parts ...