Peak Testing Commands
These commands are used to read the maximum/minimum value after startup test. Peak value recording is automatically
cleared when the Peak test is started.
6.1 PEAK[:STATe] <bool>
This command is used to set the peak test status.
Command Syntax
PEAK[:STATe] <bool>
Example
Peak:ON
6.2 PEAK CLEar
This command is used to clear the peak record.
Command Syntax
PEAK CLEar
Example
Peak CLE
6.3 PEAK:VOLTage:MAXimum?
This command is used to query the maximum voltage.
Command Syntax
PEAK:VOLTage:MAXimum?
Example
PEAK:VOLT:MAX?
Query Response
<NR2>
6.4 PEAK:VOLTage:MINimum?
This command is used to query the minimum voltage.
Command Syntax
PEAK:VOLTage:MINimum?
54
Peak Testing Commands
Example
PEAK:VOLT:MIN?
Query Response
<NR2>
6.5 PEAK:CURRent:MAXimum?
This command is used to query the maximum current.