0 OR OFF: The value of status byte enable register, operater event enable register, query event enable register and
standard event enable register will be stored in the none-volatile storage, which will be recalled when power on.
Command Syntax
*PSC <bool>
Parameters
0- 1- ON- OFF
Query Syntax
*PSC?
Query Response
0- 1
1.7 *RCL
This command restores the electronic load to a state that was previously stored in memory with a *SAV command to the
specified location. All states are recalled with the following exceptions:
CAL:STATe is set to OFF
The trigger system is set to the Idle state by an implied ABORt command (this cancels any uncompleted trigger actions)
NOTE: The device state stored in location 0 is automatically recalled at power turn-on.
Command Syntax
*RCL <NRf>
Parameters
0 to 9
Example
*RCL 3
Related Commands
*PSC *RST *SAV
1.8 *RST
This command reset the electronic load to the factory-defined states.
Command Syntax
*RST
Parameters
None
1.9 *SAV
This command stores the present state of the electronic load to a specified location in memory. Up to 100 states can be
stored. If a particular state is desired at power-on, it should be stored in location 0. It then will be recalled at power-on
if the power-on state is set to RCL0. Use *RCL to retrieve instrument states.
Command Syntax
*SAV <NRf>
Parameters
0 - 99
Example
*SAV 3
Related Commands
*PSC *RST *RCL
1.10 *SRE
This command sets the condition of the Service Request Enable Register. This register determines which bits from the
Status Byte Register (see *STB for its bit configuration) are allowed to set the Master Status Summary (MSS) bit
and the Request for Service (RQS) summary bit. A 1 in any Service Request Enable Register bit position enables the
corresponding Status Byte Register bit and all such enabled bits then are logically ORed to cause Bit 6 of the Status
Byte Register to be set.