37
Command Syntax
TIMing:LOAD:MODE <mode>
Parameters
CURR- VOLT- POW- RES
Example
TIM:LOAD:MODE CURR
Query Syntax
TIMing:LOAD:MODE?
Query Response
<mode>
Related Commands
If is executed, the setting of this command does not take effect.
10.4 TIM:LOAD:SETT OFF
10.5 TIMing:LOAD:VALue
This command is used to set the load parameters for the Timing test.
Command Syntax
TIMing:LOAD:VALue <Nrf+>
Parameters
A / V / W / , depending on
TIMing:LOAD:MODE
Example
TIM:LOAD:VAL 1
Query Syntax
TIMing:LOAD:VALue?
Query Response
<NR2>
Related Commands
If
TIM:LOAD:SETT OFF
is executed, the setting of this command does not take effect.
10.6 TIMing:TSTart:SOURce
This command sets the trigger source for starting the test.
Command Syntax
TIMing:TSTart:SOURce <source>
Parameters
VOLT- CURR- EXT
Example
TIM:TST:SOUR VOLT
Query Syntax
TIMing:TSTart:SOURce?
Query Response
<source>
10.7 TIMing:TSTart:EDGE
This command is used to set the trigger edge of the start test.
Command Syntax
TIMing:TSTart:EDGE <edge>
Parameters
RISE- FALL
Example
TIM:TST:EDGE RISE
Query Syntax
TIMing:TSTart:EDGE?
Query Response
<edge>