CHAPTER 7. INTEGRAL OPERATION FUNCTION
38
–
Average power integral (WPAV):
±
99999 Mw
•
Restricted implementation
Settings of the following functions cannot be changed during integral operation.
NOTE In addition to the above items, other settings (such as self-testing and date/time setting) or operations with
influence on the integral cannot be executed; otherwise, errors may be caused.
7.5
Setting of Integral Measurement Configuration
You can set the start mode, stop mode, automatic zero setting, automatic calibration, watt-hour integration,
current integration and other parameters of integral measurement. Specific steps are as follows:
Operation steps
1.
Press “Integ” to enter the integral measurement inter
face.
2.
Press the soft key corresponding to the “SETUP” parameter in the integral measurement interface to enter
the integral parameter configuration interface. Press the “ ” button to select the required parameter, as
shown in the figure below.
7.5.1
Parameter descriptions:
Start set in the MANUAL mode or TIME mode via the right soft key.
MANUAL press the “START” soft key in the integral measurement interface in the manual start mode to trigger the
manual start of the integral function.
TIME: the date on which the integral mode is started can be set in the real-time start mode. Integration will be
started on the set date. When the time/date combination is set in the real-time start mode, integration will
not be started before the current time and date. Integration will be started when the screen is updated once
at least before the start time.
Stop: set in the MANUAL mode, TIME mode and TINTerval mode via the right soft key.
MANUAL: press the “STOP” soft key in theintegral measurement interface in the manual st
op mode tomanually
stop integration. When the integration time reaches the maximum integration time (10,000 hours) or the
integral value reaches the maximum/minimum display value, integration will be stopped, and the current
integral time and integral value will be kept.
TIME The date for stop of the integral mode can be set in the real-time stop mode. Integration will be stopped on
the set date. When it reaches the set time or the integral value reaches the maximum/minimum display
integral value, integration will be stopped, and the current integral time and integral value will be kept.
TINTerval the integral measurement time can be set in the set-time stop mode. When it reaches the set ending
date and time or the integral value reaches the maximum/minimum display integral value, integration will be
stopped, and the current integral time and integral value will be kept.
Auto Clear used for enabling (ON) or disabling (OFF) the function of automatic zero clearing of the integral. On:
integration is restarted when the START button is pressed to start integration. OFF: integration is restarted
based on the integral value at the end of previous integration when the START button is pressed to start
integration.