CHAPTER 3. FRONT PANEL OPERATION
17
3.5
Averaging Function
The user can set the averaging function via this menu. When the input signal frequency is low, the value is
displayed unstably and cannot be read easily. In this case, the averaging function can be enabled to calculate and
display the average value of several measurements.
1.
Select MENU
⇒
SET UP
⇒
AVERAG SET to enter the configuration page of the averaging function.
2.
Press
to select the parameter to be configured (blue font background), and then press the soft key
corresponding to the parameter on the right to set the required value, as shown in the Figure
3.4
.
3.
Press the Enter button to confirm the setting.
Figure 3.4: Averaging Settings
Character
Function description
State
Set the status of the averaging function. When “ON” is selected, the averaging
function is enabled. When “OFF” is selected, the averaging function is disabled.
Mode
Set the mode of the averaging function. EXP: index averaging, often used for analysis
of the non-stationary process. LINE: linear averaging, often used for measurement
and analysis of the stationary random process. The deviation relative to the standard
can be reduced by increasing the averaging times.
Type
Set the linear averaging type. MOVING: moving averaging REPEAT: repeated
averaging
Count
Set the times of the averaging function. If the mode of the averaging function is set
as EXP (index averaging), the attenuation constant can be set. If the mode of the
averaging function is set as LINE (linear averaging), the averaging times can be set.
Table 3.4: Average Settings Menu
3.5.1
Index averaging
D
n
The value displayed after the nth index averaging
(D1, the value displayed after the first averaging, is equal to M1)
D
n
−
1
The value displayed after the (n-1)th index averaging