Chapter 8
Remote Operation
There are three types of communication interfaces available: USB ,Ethernet, and RS232.You can choose any one of
them to communicate with a PC.
8.1
RS-232 Interface
RS232 interface:use a cable with two COM interface (DB9) to connect power meter and PC. It can be activated by
menu key on the front panel.All SCPI commands are available through RS-232 programming.
•
RS-232 data format The RS-232 data is a 10-bit word with one start bit and one stop bit. The number of start
and stop bits is not programmable.
•
Baudrate Available baudrate:4800 9600 19200 38400 57600 115200
•
RS-232 connection The RS-232 serial port can be connected to the serial port of a controller (i.e., personal
computer) using a straight through RS-232 cable terminated with DB-9 connectiors. Do not use a null
modem cable. Table
8.1
shows the pinout for the connector.
If your computer uses a DB-25connector for the RS-232interface,you will need a cable or adapter with a DB-25
connector on one end and a DB-9 connector on the other,wired straight through(not null modem).
RS-232
8.1.1
RS-232 troubleshooting
If you are having trouble communicating over the RS-232 interface,check the following: The computer and the
power meter must be configured for the same baudrate, parity, number of data bits,and flow control options.
Note that the eledtronic load is configured for 1 start bit and 1stop bit (these values are fixed). The correct
interface
Figure 8.1: RS-232 DE-9 Connector
41
CHAPTER 8. REMOTE OPERATION
42
Pin #
Signal
1
NC