CHAPTER 10. ROUTINE MAINTENANCE
49
Start time is less than
current
Error description The start time is less than the current time.
Possible cause The integral setting is incorrect.
Disposal Reset the integral start time.
End time is less than
current
Error description The ending time is less than the current time.
Possible cause The integral setting is incorrect.
Disposal Reset the integral ending time.
Timer must be larger than
zero.
Error description The set time must be larger than zero.
Possible cause The set integral time is not correct.
Disposal Reset the set integral time.
Harmonic open fail
Error description The harmonic function cannot be enabled.
Possible cause Communication abnormality Disposal
Check the communication cable.
Integ open fail
Error description The integral function cannot be enabled.
Possible cause Communication abnormality Disposal
Check the communication cable.
Integ start fail
Error description The integral function cannot be started.
Possible cause Communication abnormality Disposal
Check the communication cable.