92
prompt message “No external clock source!” would be displayed and the clock source
would be switched to “Internal”.
4.
The unit can also output a 10 MHz clock signal when the 10MOut is enabled. If the
output of the clock signal is enabled, then the [10MHz In/Out] connector in the rear
panel will function as an clock signal output only.
Sync methods for two or more instruments:
Synchronization between two instruments
Connect the [10MHz In/Out] connector of generator A (using internal clock) to the [10MHz
In/Out] connector of generator B (using external clock) and set the output frequencies of A and
B as a same value to achieve synchronization.
Synchronization among multiple instruments
Divide the 10 MHz clock source of a generator (using internal clock) into multiple channels and
connect them to the [10MHz In/Out] connectors of other generators (using External clock).
Then set the output frequencies of all the generators to the same value to achieve
synchronization.
4.9
Mode
The 4050B series has two modes of operation: Phase-Locked and Independent.
To select a mode of operation, follow the steps below:
1.
Press the Utility button.
2.
Press the Page 1/2 softkey.
3.
Press the Mode softkey to select a mode of operation.
Phase locked:
When changing the frequency, the DDS of both channels are reset and the phase
deviation between CH1 and CH2 is maintained.