BT-710 Operation manual
P/N : 710-ENG-OPM-EUR-R02
Bistos Co., Ltd.
2020.03
24
7.3 Alarm mode
When an alarm occurs, the monitor uses the following audible or visual alarm to prompt the
user:
Visual alarm
Audible alarm
Alarm info
Parameter flashing
Of which, the visual alarm, audible alarm, and alarm information distinguish the alarm levels
in a different manner respectively.
7.3.1 Visual alarm
When an alarm occurs, the alarm indicator will flash in different colors and frequencies to
prompt the alarm priority.
High priority alarm: Red
Medium and Low priority alarm: Yellow
7.3.2 Audible alarm
An audible alarm is that the monitor prompts the alarm priorities with alarm tone character-
istics when an alarm occurs.
Audible alarm pattern:
High priority alarm: Beep-beep-beep--beep-beep----beep-beep-beep--beep-beep
Medium priority alarm: Beep-beep-beep
Low priority alarm: Beep-
7.3.3 Alarm information
Alarm information displayed on the physiological or technical alarm area of the monitor
indicates the corresponding alarm information when an alarm occurs. The system will
distinguish the alarm priority with different background colors:
High priority alarm: Red
Medium priority alarm: Yellow
Low priority alarm: Yellow
The following flags in front of physiological alarms are used to distinguish the alarm priori-
ties.
High priority alarm: ***
Medium priority alarm: **
Low priority alarm: *
7.3.4 Parameter flashing
When the physiological parameter values in the parameter area will flash once per second,
and the indicator for the upper limit and lower limit of the parameter area will also flash at
the same frequency, it indicating that the parameter exceeds the upper limit or lower limit.