BT-710 Operation manual
P/N : 710-ENG-OPM-EUR-R02
Bistos Co., Ltd.
2020.03
35
13.2 Recommended separation distances between portable and mobile RF communica-
tions equipment and BT-710
The BT-710 is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the BT-710 can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmit-
ters) and the BT-710 as recommended below, according to the maximum output power of the communi-
cations equipment.
Rated maxi-
mum output
power of
transmitter
[W]
Separation distance according to frequency of transmitter
[m]
150 kHz to 80 MHz
d
=
3.5
�𝑝𝑝
80 MHz to 800 MHz
d
=
3.5
�𝑝𝑝
800 MHz to 2.5 GHz
d
=
[
7
3
]
�𝑝𝑝
0.01
0.35
0.35
0.23
0.1
1.11
1.11
0.74
1
3.5
3.5
2.34
10
11.07
11.07
7.38
100
35
35
23.24
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the trans-
mitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1) At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2) These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
13.3 Electromagnetic immunity
The BT-710 is intended for use in the electromagnetic environment specified below.
The customer or the user of the BT-710 should assure that it is used in such an environment.
Immunity test
IEC 60601 Test
level
Compliance level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-
2:2009
±8 kV Contact
±15 kV air
±8 kV Contact
±15 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the
relative humidity should be at least
30 %.
Electrical fast
transient/burst
IEC 61000-4-
4:2004
±2 kV for power
supply lines
±1 kV for in-
put/output lines
(>3m)
±2 kV for power
supply lines
±1 kV for in-
put/output lines
(>3m)
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-
5:2006
±1 kV differential
mode
±2 kV common
mode
±1 kV differential
mode
±2 kV common
mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips,
short interrup-
tions and
< 5 %
U
т
(> 95 %
dip in
U
т)
for 0.5
cycles
< 5 %
U
т
(> 95 %
dip in
U
т)
for 0.5
cycle
Mains power quality should be
that of a typical commercial or
hospital environment. If the user