BT-710 Operation manual
P/N : 710-ENG-OPM-EUR-R02
Bistos Co., Ltd.
2020.03
13
Description
1
Current alarm message. When an alarm occurs, this area will displayed yellow or
red depending on the alarm type.
2
SpO
2
value. Display the measured SpO
2
value.
3
SpO
2
upper alarm limit. Display the user set upper alarm limit
4
SpO2 lower alarm limit. Display the user set lower alarm limit
5
SpO
2
waveform. Display the measured SpO
2
waveform. The waveform is not
normalized.
6
Pulse rate value. Display the measured pulse rate per minute.
7
Sleep mode. Touch this area makes the pulse oximeter to enter the sleep mode.
To exit "sleep mode", press [Power] or [Alarm reset] or [Setting]
8
Perfusion Index. Display the measured perfusion index.
9
Pulse rhythm strength.
10
Battery Status
11
Patient type.
12
The unit of SpO
2
.
1.7 Essential performance
This device Pulse oximeter provides patient vital signs such as pulse rate, blood oxygen
saturation and perfusion index by placing the sensors to the appropriate site of patient. The
device is composed with display, control circuit and panel, and input part for SpO
2
sensors. It
detects SpO2 and PR using specific sensors. The detected analog signal amplified and con-
verted to digital. This converted data feed to the CPU and converted to the display format as
number and waveform. This device is incorporated with alarm system. The alarm generated
when the detected signal range is beyond the user set alarm limits.
2 Preparing for operations
2.1 Installation
To ensure normal working of the pulse oximeter, read this chapter before use, and install as
required.
WARNING
•
All analog and digital devices connected to the pulse oximeter must be certi-
fied by IEC standards (e.g. IEC 60950 Data processing equipment standard and
IEC 60601-1 Medical equipment standard). Furthermore, all configurations
shall comply with valid version of IEC 60601-1 standard. The personnel con-
necting additional devices to the input / output signal ports are responsible for
the compliance with IEC 60601-1 standard. If there is any question, please con-
tact Bistos.