BT-710 Operation manual
P/N : 710-ENG-OPM-EUR-R02
Bistos Co., Ltd.
2020.03
29
Hydrogen peroxide (3%)
Ethanol (70%)
Isopropanol (70%)
Before cleaning:
Turn off the pulse oximeter, disconnect the power cord.
Use a soft cotton ball to adsorb appropriate amount of cleaning agent and wipe the
display screen.
Use a soft lint-free cloth to adsorb appropriate amount of cleaning agent and wipe the
surface of the device.
If necessary, use a clean, dry, lint-free cloth to remove any excess detergent.
Dry the device naturally in a ventilated cool environment.
WARNING
•
Before cleaning the pulse oximeter or sensor, turn off the power and discon-
nect the DC power.
•
The pulse oximeter should be kept clean. It is recommended to regularly clean
the enclosure surface and the display screen. Cleaning the enclosure with non-
etching cleaner such as soap and water.
CAUTION
•
To avoid damaging the pulse oximeter:
•
Do not use strong solvents such as acetone.
•
Most cleaners must be diluted before use. Diluting should be according to the
manufacturer's instructions.
•
Do not use abrasive materials (such as steel wool).
•
Do not allow any liquid entering into the enclosure, and never immerse any
part of the device into liquid.
•
Do not leave any cleaning solution on the surface of any part of the device.
NOTE
•
Wipe the pulse oximeter and sensor surface with medical alcohol, dry it
naturally or with clean, dry, lint-free cloth.
•
Bistos is not liable for effectiveness of using these chemicals for infectious
disease control. Please consult the infectious disease control officers or ex-
perts of the hospital for advice.
9.3 Disinfection
In order to avoid damage to the product, we recommend that the product is disinfected only
when it is deemed necessary by the hospital maintenance procedures. We also recommend