BT-710 Operation manual
P/N : 710-ENG-OPM-EUR-R02
Bistos Co., Ltd.
2020.03
19
NOTE
•
Do not put the oxygen probe and blood pressure cuff on the same limb, be-
cause blood flow occlusion during blood pressure measurement will affect the
SpO
2
readings.
•
The pulse oximeter cannot be used to verify the accuracy of SpO
2
probe and
SpO
2
equipment.
5.3 Monitoring steps
(1)
Act the appropriate SpO
2
sensor according to the patient.
(2)
Turn on the pulse oximeter, and connect the SpO
2
sensor probe to the pulse oximeter.
(3)
Clean the measurement site, such as finger with nail polish.
(4)
Put the SpO
2
sensor probe on the patient’s finger.
(5)
Select the appropriate alarm settings.
(6)
Start monitoring.
NOTE
•
When you turn on the pulse oximeter, plug in SpO
2
probe and connect pa-
tient’s finger (or toe), monitor displays SpO
2
wave, “SpO
2
Pulse Search” dis-
played in the technical alarm area until the monitor measured SpO
2
value and
pulse rate. “SpO
2
Search Timeout” displayed in the technical alarm area until
the pulse oximeter measured pulse rate. Check the sensor mounting position,
whether the sensor is damaged or sensor type. Reconnect the sensor or use
new sensor.
•
The alarm message for signal inadequacy such as “SpO2 No Sensor connect-
ed”, “SpO2 Sensor Off”, “SpO2 Search Timeout”, “SpO2 Signal Unstable”,
“SpO2 Sensor Error” and low priority alarm and indication for signal instability
like “---” are meaning that the value of SpO2 or PR might be inaccurate.
5.4 Setting SpO
2
Select the main screen to enter the SpO2 set interface
Select “Wave Speed” and set wave speed to “12.5mm/s” or “25mm/s”. The faster speed,
the smoother wave.