BT-550 Operation manual
7
P/N : 550-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2018.03
normal operation in the configuration in which it is to be used can be achieved.
For proper operation of the infant warmer, use only skin temperature sensor
from Bistos Co. Ltd. Using other sensor could result in personal injury or
equipment damage.
Never place the skin temperature sensor under the infant’s body or use it
rectally. Personal injury could occur.
The skin temperature sensor must be in direct contact with the skin to provide
accurate measurement of the infant’s skin temperature. Failure to maintain
direct skin contact can result in overheating. Routinely check the infant’s
condition for correct sensor attachment, and feel the infant’s skin for signs of
overheating.
Do not use in the presence of flammable anesthetics. Personal injury or
equipment damage could occur.
Keep matches, and all other sources of ignition, out of the room in which the
infant warmer is located. Textiles, oils, and other combustibles are easily ignited
and burn with great intensity in air enriched with oxygen. Personal injury or
equipment damage could occur.
Small quantities of flammable agents, such as ethyl and alcohol, left on the
infant warmer may cause a fire in connection with oxygen. Personal injury or
equipment damage could occur.
A fire and explosion hazard exists when performing cleaning or maintenance
procedures in an oxygen-enriched environment.
To avoid overheating the infant due to direct radiation, do not position the
infant warmer in direct sunlight or under other sources of radiant heat.
To prevent accidental disconnection when adjust the height, ensure that all
patient leads, infusion lines, and ventilator tubing have sufficient excess length.
For infant’s safety, do not leave the infant unattended. Personal injury could
occur.
Never place objects taller than the top of the wheel casters beneath the infant
warmer stand. Personal injury of equipment damage could occur.
For optimum stability, always lock all stand wheels. Do not leave the device
unattended when parking on an incline. Failure to do so could result in personal
injury or equipment damage.
Prior to placing the infant in the infant warmer, pre-warm the BT-550 to the
temperature prescribed by the attending physician, or according to nursing
protocol.
Only one monitor shelf should be used per infant warmer. When using the
monitor shelf, always place the monitor in the center of the shelf, ensure that