BT-550 Operation manual
48
P/N : 550-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2018.03
4
Maintenance and cleaning
4.1 Service policy
The repair of infant warmer should be performed qualified service personnel. Contact to Customer
Service Department of Bistos when the device fault or malfunctioning.
4.2 Maintenance
It is recommended to have a safety inspection every year regularly for the safe use of BT-550. Refer to
the service manual for the inspection items and method.
WARNING
Do not remove the covers of BT-550 yourself to avoid damage to the infant warmer
and unexpected electrical shock. Only qualified Bistos service engineer must repair
or replace components.
The life time of heater is about 5 000 hours. After this time the heater should be
changed.
4.3 Cleaning
Follow the instruction prior to clean the infant warmer.
Turn off the BT-550 and detach the power cord.
Detach all accessories from BT-550.
Cool down the BT-550 for about 30 minutes.
Use a clean gauze pad or lint-free cloth, lightly moistened with a water or mild detergent, to
wipe the surface of BT-550.
WARNING
Do not use the conductive or flammable cleaning agent to avoid electric shock, fire
or explosion.
Do not use spray type cleaning agent.
Do not scrub the display with rough or sharp objects.
Do not use steam to clean the infant warmer. Excessive steam can damage the
infant warmer.
Avoid using alcohol and strong detergent to bassinet made of acrylic.
Ensure no part of the warmer or accessories are immersed in any cleaning agents.