Turn the power off after using the device. When the device expected do not used for a while,
then unplug the power cord from socket outlet.
2.8 SpO
2
Monitor(option)
This symbol means that the Masimo product inside.
BT-550 uses the Masimo SpO2 oximetry to monitor the oxygen saturation in the blood. You can find
the detailed Masimo product information below.
Patent information for Masimo product can be found at www.masimo.com/patents.htm.
“Possession or purchase of this device does not convey any express or implied license to use the device
with unauthorized sensors or cables which would, alone or in combination with this device, fall within
the scope of one or more the patents relating to this device”
WARNING
SpO
2
monitor of BT-550 must be operated by qualified personnel only.
Do not use the MS board pulse oximeter in the presence of flammable anesthetics
or other flammable substance in combination with air, oxygen-enriched
environments, or nitrous oxide.
A pulse oximeter should NOT be used as an apnea monitor.
Pulse rate measurement is based on the optical detection of a peripheral flow
pulse and therefore may not detect certain arrhythmias. The pulse oximeter
should not be used as a replacement or substitute for ECG based arrhythmia
analysis.
A pulse oximeter should be considered an early warning device. As a trend
towards patient deoxygenation is indicated, blood samples should be analyzed by
a laboratory co-oximeter to completely understand the patient’s condition.
If an alarm condition (other than exceptions listed herein) occurs while the alarm
silence period is set to off, the only alarm indications will be visual displays and
symbols related to the alarm condition.
The MS board pulse oximeter is to be operated by qualifier personnel only. These
manual, accessory directions for use, all precautionary information, and
specifications should be read before use.
Do not remove the monitor cover except to replace the battery. An operator may
only perform maintenance procedures specifically described in this manual. Refer
servicing to Masimo in repair of this equipment.
Measure the oximeter’s leakage current whenever an external device is connected