BT-450 Operation manual
P/N:450-ENG-OPM-EXP-R01
Bistos Co., Ltd.
2021.06
11
WARNING
•
If the high temperature icon appears, turn off the control box and check
the patient’s body temperature.
•
If the low battery information signal is on, connect the adapter to the
device.
3.
How to use
3.1 Preparation before use
①
Check the arrow symbols on the pad connector and the control box and then combine them.
NOTE
When disconnecting
: Pull the metal part toward the pad to disengage it.
No
Name
Description
①
High temperature icon
If the contact part temperature of the pad is high, the pad is
automatically turned off before the temperature exceeds 40 °C.
The high temperature icon appears on the display.
②
Pad connection icon
If the communication between the control box and the pad is
not normal, the pad connection icon appears on the display.
③
Light intensity
Display the current set light intensity (High, Low, OFF).
④
Battery icon
Display the current battery level.
⑤
Charging icon
The charging icon appears when the adaptor charges the device.
⑥
Operating Time
Display the elapsed time from the pad was on.
⑦
Remaining Time
Display the remaining time set in the timer mode.
⑧
Total Usage Time
Display the total usage time from the pad was on for the first
time.
⑨
Firmware version
Display the currently downloaded firmware version.