BT-450 Operation manual
P/N:450-ENG-OPM-EXP-R01
Bistos Co., Ltd.
2021.06
17
7.2 Cleaning
(1) Wash
-
Before cleaning, turn off the device and disconnect the pad connector.
-
Remove residues on the surface of the control box and pad with a soft brush or soft clean-
cloth wet with a proper amount of clear water.
(2) Disinfection
-
Before disinfection, turn off the device and disconnect the pad connector.
-
Clean the surface of the control box and pad with a soft brush or soft clean-cloth wet with a
proper amount of 70 ~ 90 % alcohol (Ethanol alcohol or isopropyl alcohol).
-
Always follow the hospital’s hygiene regulations when handling devices contaminated with
bodily fluids or other substances.
WARNING
Disconnect the AC power cable before cleaning.
Turn off the device and disconnect the pad connector before cleaning.
CAUTION
Clean the pads before use on a new patient.
8.
Troubleshooting
Problem
Solution
The battery cannot be
charged.
•
Check if the power adapter is properly connected.
•
If the charging problem persists, replace the power adapter with
a new one.
•
If the charging problem persists, the battery needs to be
replaced. Contact Bistos Service Center.
The power does not turn
on.
•
Check the battery charging status.
•
Inspect the power button and the outside of the control box for
foreign substances.
The control box gets wet.
•
Remove the plug of the power adapter from the socket.
•
Turn off the power of the control box.
•
Wipe the body with a dry cloth and store it in a warm dry place
for at least 12 hours.
LED does not light up.
•
Check if the pad is properly connected to the control box.
•
If parts of the LED module do not light on, the pad needs to be
replaced.
The pad connection icon
still appears.
•
Unplug the pad connector and reconnect it to the control box.
•
If the pad connection icon still appears, the pad needs to be
replaced. Contact Bistos Service Center.
※
If the problem is not solved or have additional questions, contact Bistos Service Center.
9.
Manufacturer’s declaration on EMC
The Infant Phototherapy Equipment needs special precautions regarding EMC (Electromagnetic
compatibility) and needs to be used according to the EMC information provided in this user manual.
Wireless communications equipment such as wireless home network devices, mobile phones, cordless
telephones and their base stations, walkie-talkies can affect the Infant Phototherapy Equipment and should
be kept at least 1 m away from the equipment. And it does not suitable for use in an MRI environment.