BT-450 Operation manual
P/N:450-ENG-OPM-EXP-R01
Bistos Co., Ltd.
2021.06
13
WARNING
•
To minimize the heat between the pad and the patient, the patient
should not be wrapped in a thick blanket or wrapped too tightly.
•
The patient’s body temperature may rise if the patient and pads are
wrapped in a material that does not allow heat to escape, such as a thick
blanket or clothes.
•
Do not use the device without a disposable cover.
•
Do not use the device with other thermotherapy devices that may affect
the patient’s body temperature such as incubators, heaters, mattresses,
heating blankets, etc.
3.3 Product Operation
3.3.1 Operation mode
Mode
Description
Normal
Mode
The pad is on when the timer indicates 0 then the device operates in the normal
mode.
Timer
Mode
The pad is on when the timer indicates higher than 0minutes; the device operates in
Timer Mode. In timer mode, the pad operates as long as the set time. When the set
time reaches 0 minutes, the pad turns off.
And the maximum setting time is 24hours.
The pad automatically turns off after 24hours.
3.3.2 How to operate
(1) Power ON/OFF
1) Press the power button ( ) for more than 1 second to turn the power ON or OFF.
2) When booting, the total usage time and F/W version are displayed at the bottom of the LCD display.
< Boot screen >
< Basic screen >
(2) Mode setting
1) Normal mode
①
Use the 30 minutes decrease
button( ) to set the Timer 0
minutes.
②
Set the light intensity to High or Low by
using the light intensity control
button( ).
③
In the normal mode, the power is not
automatically turned off, so the user must
manually turn off the power.