BT-450 Operation manual
P/N:450-ENG-OPM-EXP-R01
Bistos Co., Ltd.
2021.06
7
Eye protection: Do not look directly into the LED. During treatment, always protect your baby’s
eyes with an eye patch or protective equipment. Periodically, check the hospital or treatment
protocol and makes sure that the baby’s eyes are protected from contamination. Patients near
the light should use protective pads or equipment to protect their eyes.
Sensitive people may develop headaches, nausea, or mild dizziness if left in the irradiated area
for too long. Wearing yellow lens glasses can reduce the potential impact.
Bilirubin Photo isomers may have toxic effects.
Water balance: Water balance may be disturbed for some patients.
Photosensitive Drugs: Irradiation may reduce the effectiveness of light-sensitive drugs. Do not
store the medication near light irradiation.
Flammable gas: Do not use the device near combustible gases. (eg oxygen, nitrogen dioxide, or
other anesthetics)
Power off: When cleaning the light source, always turn off the power and remove the power
cable.
Even an adult may be affected by staying in the light for a long time.
Do not use flammable solution directly on the LED lamp. The LED lamp may be damaged or its
function may deteriorate. For cleaning or maintenance, follow the instructions described in
section 7 of this manual.
Use of other accessories not provided by Bistos Co., Ltd. may degrade the safety and
performance of the device, so be sure to use the designated accessories.
Using a reflective film can cause an increase in body temperature when the film affects a type of
phototherapy radiation.
Blue light might interfere with the clinical management of a patient with cyanosis.
Infants with increased direct reactions or cholestatic jaundice after receiving phototherapy may
develop the bronze-baby syndrome.
Do not use any damaged parts including the power adapter.
Before using the device, visually check each part for cracks, folds, tears, discoloration, or
deterioration. If damage to the device is observed, discontinue until the part is replaced.
Do not use the BT-450 for neonatal jaundice treatment near or in combination with other
equipment. If you want to use adjacent to or overlapping with devices, you must confirm the
normal operation or BT-450.
Portable RF communication equipment (including peripheral devices such as antenna cables and
external antennas) should not be used closer than 30cm (12 inches) to any part of the BT-450,
including cables specified by the manufacturer. Otherwise, the performance of this device may
be degraded. Wireless home network devices, mobile phones, wireless phones, and wireless
communications equipment such as base stations and radios can affect the device and should be
used at a distance of 1 m or more.
This device requires special precautions regarding EMC and must be serviced according to the
EMC information provided in this manual.
Excessive electromagnetic interference may cause the device to generate excessively high or low
light levels temporarily. Use the device in an appropriate environment stated in this manual.
Components other than those approved for use with this device may increase or decrease
immunity.
Do not open the pad cable where there the power is connected. There is a risk of electric shock.
Do not disassemble the device. The battery can be replaced by disassembling the device by an
authorized service person and must be replaced with a battery of the same model and
specification.
Incorrect battery replacement may cause high temperature, fire, or explosion.
If the device is not used for a long time, the battery must be removed and stored. Charge the
battery and store it at room temperature with low humidity.
Use only the AC/DC power adapter supplied or designated by Bistos. Use of adapters other than
those provided or specified by the manufacturer may increase electromagnetic emissions or
reduce the electromagnetic immunity of the device, resulting in improper operation.