SignalHawk™ PC and Rack Mount Models
91
Detection Modes
Sample, +peak, –peak, average, video average
Trigger Modes
Single, free run
Trigger Sources
Internal, external TTL, internal video
External Trigger Types
Rising edge, falling edge, any edge, trigger on HIGH,
trigger on LOW
External Trigger Level
TTL Levels
External Trigger Delay
User-settable, 100
μ
s to 1000 ms after trigger received
before starting sweep
External Trigger Connector
BNC(F)
Markers
6 Markers; Modes: On/Off, Standard, Marker to Max
Peak, Marker to Next Peak, Marker Freq to Center,
Marker Ampt to Ref Level, Marker Delta, Marker Type
Line/Icon, Marker Noise, All Marker to Peaks, All
Markers Off. Also, Marker to Max/Min Peak
General Specifications
Model
Name: Hand-Held, PC SignalHawk
Model: SH-36S-PC
Description: PC-based Spectrum Analyzer
Housing
Aluminum Case, Plastic corner protectors
Power Indicator
Green LED
Charge Indicator
Green/Amber LED – Amber: low battery or charging,
Green: fully charged.
Connectors
N(F) for RF input
2.1 mm DC jack for AC adapter/charger external power
USB Type B for PC connection
BNC(F) for external trigger
RF Input Impedance
50 ohms, nominal
RF Input VSWR
1.8:1 typ, 2.0:1 max
(internal attenuator = 10dB or greater)
Max DC Input
± 50 V
Max RF Input
+20 dBm; damage level is +30dBm for 30 sec.
AC Adapter/Charger
External DC power supply, 12 VDC, 2 A, 2.1mm pin
connector
Internal Battery
Rechargeable, field replaceable, lithium-ion battery.
5.5 hours continuous operation, 25°C.
Warm-Up
Specifications apply after a 30 minute warm-up period at
ambient temperature. Typical values are provided for
reference and are not guaranteed.
Operating Temperature
0 to 50 °C (MIL-PRF-28800F, Class 3)
Measurements
Содержание SignalHawk SH-36S-PC
Страница 26: ...SignalHawk PC and Rack Mount Models 11 Figure 7 Select DSP Device List ...
Страница 67: ...Measurements 52 Figure 41 Example Channel Power Integration Bandwidth STEP 2 STEP 3 ...
Страница 75: ...Measurements 60 Figure 49 Example Carrier to Interference Ratio Carrier Off STEP 4 Measurement Settings ...
Страница 116: ...SignalHawk PC and Rack Mount Models 101 Figure 75 Menu Map BW Sweep Menu All Measurements ...
Страница 117: ...Menu Maps 102 Figure 76 Menu Map Amplitude All Measurements ...
Страница 118: ...SignalHawk PC and Rack Mount Models 103 Figure 77 Menu Map Measurement All Measurements ...
Страница 119: ...Menu Maps 104 Figure 78 Menu Map Mark Limit Menu All Measurements ...
Страница 120: ...SignalHawk PC and Rack Mount Models 105 Figure 79 Menu Map File Help All Measurements ...
Страница 121: ...Menu Maps 106 Setup Function Menu Maps Figure 80 Map Setup Mode Main Screen ...