Medical MRI3T Power Measurement Kit
53
Sensor (Model 5014M)
Elements
Cables
RF Power Handling
As specified under General Specification
Frequency Range
50 – 130 MHz, with additional calibration at
specified freq and power levels
Cal freq & nominal power level
127.72 MHz
63.86 MHz
51.70 MHz
35 kW, 25 kW, 4.5 kW, 3.3 kW
16 kW, 2 kW
8 kW, 4 kW, 2.5 kW
Settling Time
< 3 sec.
Accuracy
General
@ cal freq & power
± 8%
± 7½%
Elements
*
* RFL element power must be 1/10 of FWD power.
Can use Bird 43 or APM Elements.
43 elements are standard
Connectors
Input
Output
7/16 (M)
7/16 (F)
Insertion VSWR, Max
1.05:1
Insertion Loss, Max
0.1 dB
Dynamic Range
20 dB
Dimensions, Nominal
2.5” x 5.0” x 2.0” (64 x 127 x 51 mm)
Weight, Nominal
1.25 lb. (0.6 kg)
Dynamic Range
20 dB
Labeling, Element
Label lists max power (W)
Connectors
7/16 (M) to 7/16 (M)
Material,
7/16 cable
RG393