Instrument Qualification Procedures
Luminescence Test
Luminescence Testing
, starting on page 81, describes the test method, lists the Gen5
protocol parameters, explains how to analyze the test results, and provides
troubleshooting tips in the event of test failure.
Required Materials
l
Harta Luminometer Reference Microplate, PN 8030015 (which includes microplate
carrier adapter PN 8032028)
l
Gen5 protocol
Synergy LX_LumTest_Harta.prt
, described on page 82
l
LUM filter cube (EX: Plug, EM: Hole, Mirror: Empty)
Test Procedure
1. Turn on the Harta reference plate using the I/O switch on the back of the plate.
2. Check the plate’s battery by pressing the test button on the back of the plate and
ensuring that the test light turns on. If the light does not turn on, replace the battery.
Synergy LX User Manual
1501000N Revision A