| Appendix
A
:
Specifications
Linearity
By liquid dilution.
96-well plate, normal read speed:
0.000 to 2.000 OD: ±1.0% ±0.010 OD, delay after plate
movement = 100 ms
2.000 to 2.500 OD: ±3.0% ±0.010 OD, delay after plate
movement = 100 ms
384-well plate, normal read speed:
0.000 to 1.500 OD: ±2.0% ±0.010 OD, delay after plate
movement = 100 ms
1.500 to 2.000 OD: ±5.0% ±0.010 OD, delay after plate
movement = 100 ms
96- and 384-well plate, sweep read
speed:
0.000 to 1.000 OD: ±1.0% ±0.010 OD
Repeatability (Standard Deviation [STD])
Tested with certified neutral density glass measured by one standard deviation (8
measurements/data point)
96-well plate, normal read speed:
0.000 to 2.000 OD: ±1.0% ±0.005 OD, delay after plate
movement = 100 ms
2.000 to 2.500 OD: ±3.0% ±0.005 OD, delay after plate
movement = 100 ms
384-well plate, normal read speed:
0.000 to 1.500 OD: ±1.0% ±0.005 OD, delay after plate
movement = 100 ms
1.500 to 2.000 OD: ±3.0% ±0.005 OD, delay after plate
movement = 100 ms
96- and 384-well plate, sweep read
speed:
0.000 to 1.000 OD: ±2.0% ±0.010 OD
Assay Validation Using a Take3 Plate
260 nm dsDNA detection limit: < 5 ng/μL
Synergy LX User Manual
1501000N Revision A
BioTek Instruments, Inc.
Содержание Synergy LX
Страница 45: ...Repackaging and Shipping Instructions 21 Synergy LX User Manual 1501000N Revision A BioTek Instruments Inc ...
Страница 46: ...22 Chapter 2 Installation Synergy LX User Manual 1501000N Revision A BioTek Instruments Inc ...
Страница 55: ...Filter Cubes 31 Synergy LX User Manual 1501000N Revision A BioTek Instruments Inc ...
Страница 138: ...114 Appendix B Sample Reports Sample System Report Synergy LX User Manual 1501000N Revision A BioTek Instruments Inc ...
Страница 139: ...Sample System Report 115 Synergy LX User Manual 1501000N Revision A BioTek Instruments Inc ...