132
| Appendix
C
:
Error Conditions
Error Code
Description
Troubleshooting
3F00
Absorbance target gain calibration is performed for the
reference channel, shooting for a target at full-scale
saturation. A signal error is flagged if the maximum gain of
254 does not result in saturation.
• Ensure that the door is
completely closed.
• Ensure that instrument
enclosure is completely
installed and secured.
• Flash lamp missing flashes
or too much flash to flash
intensity variation.
• Check for spill in chamber.
Contamination on the lower
or the upper optical
assembly in the absorbance
light path can cause this
error. A system test will
typically show elevated
gains in the 230 nm
wavelength if the
absorbance optics is
contaminated.
• Contact Technical Support.
Note: This error may not
occur during a system test
but can occur when the first
read is attempted.
3F0Yw
Yw = wavelength readset #
If the final signal for the reference channel (dark offset
subtracted from light) is less than 500 counts, or if the
measurement channel signal is less than 8000 counts, a
signal error is flagged.
Synergy LX User Manual
1501000N Revision A
BioTek Instruments, Inc.
Содержание Synergy LX
Страница 45: ...Repackaging and Shipping Instructions 21 Synergy LX User Manual 1501000N Revision A BioTek Instruments Inc ...
Страница 46: ...22 Chapter 2 Installation Synergy LX User Manual 1501000N Revision A BioTek Instruments Inc ...
Страница 55: ...Filter Cubes 31 Synergy LX User Manual 1501000N Revision A BioTek Instruments Inc ...
Страница 138: ...114 Appendix B Sample Reports Sample System Report Synergy LX User Manual 1501000N Revision A BioTek Instruments Inc ...
Страница 139: ...Sample System Report 115 Synergy LX User Manual 1501000N Revision A BioTek Instruments Inc ...