Error Codes
|
121
Error
Code
Description
Troubleshooting
A100
Task not available. A requested subroutine or module has been
requested but is not available.
If repeatable, contact
Technical Support.
A200
Code version strings for digital PCB processors do not match.
A300
The x-axis hardware module is not available.
A301
The y-axis hardware module is not available.
A302
The z-axis hardware module is not available.
A303
The order-sorting filter wheel hardware module is not
available.
A304
The monochromator hardware module is not available.
A500
EEProm memory read/write error
AB00
Multiprocessor error
AC00
External RAM error
ADXX
Stack error
AEXX
Invalid motor ramp profile
7FXX
Dual processor error
Synergy LX User Manual
1501000N Revision A
BioTek Instruments, Inc.
Содержание Synergy LX
Страница 45: ...Repackaging and Shipping Instructions 21 Synergy LX User Manual 1501000N Revision A BioTek Instruments Inc ...
Страница 46: ...22 Chapter 2 Installation Synergy LX User Manual 1501000N Revision A BioTek Instruments Inc ...
Страница 55: ...Filter Cubes 31 Synergy LX User Manual 1501000N Revision A BioTek Instruments Inc ...
Страница 138: ...114 Appendix B Sample Reports Sample System Report Synergy LX User Manual 1501000N Revision A BioTek Instruments Inc ...
Страница 139: ...Sample System Report 115 Synergy LX User Manual 1501000N Revision A BioTek Instruments Inc ...