Fluorescence Specifications
|
112
Fluorescence Specifications
Detection limit: Sodium Fluorescein in
phosphate buffered saline (PBS)
DL ≤20 pM, Excitation 485/20 nm, Emission
528/20 nm, 510 nm mirror
Detection limit: Methylumbelliferone (MUB)
in carbonate-bicarbonate buffer (CBB)
DL ≤0.16 ng/mL (0.91 nM), Excitation 360/40
nm, Emission 460/40 nm, 400 nm mirror
Minimum kinetic interval:
< 55 seconds for a 96-well microplate with the
following read settings:
Delay after plate movement: 100 ms
Measurements per data point: 10
Wavelength range:
320–700 nm (low-noise PMT)
320–850 nm (red-shifted PMT)
Luminescence Specifications
Production testing is performed using a Harta plate.
96-well plate with standard low-noise PMT
Detection Limit for ATP:
≤ 75 amol/well
Integration Time:
10 seconds
PMT Gain:
150
Blank Wells:
16
Wavelength Range:
200–700 nm
96-well plate with red-shifted PM
Detection Limit for ATP:
≤ 500 amol/well
Integration Time:
10 seconds
PMT Gain:
150
Blank Wells:
16
Wavelength Range:
200–850 nm
Synergy LX User Manual
1501000N Revision A
BioTek Instruments, Inc.
Содержание Synergy LX
Страница 45: ...Repackaging and Shipping Instructions 21 Synergy LX User Manual 1501000N Revision A BioTek Instruments Inc ...
Страница 46: ...22 Chapter 2 Installation Synergy LX User Manual 1501000N Revision A BioTek Instruments Inc ...
Страница 55: ...Filter Cubes 31 Synergy LX User Manual 1501000N Revision A BioTek Instruments Inc ...
Страница 138: ...114 Appendix B Sample Reports Sample System Report Synergy LX User Manual 1501000N Revision A BioTek Instruments Inc ...
Страница 139: ...Sample System Report 115 Synergy LX User Manual 1501000N Revision A BioTek Instruments Inc ...