3-24
Operation
•
P-ACROSS
enables a blank in every even-numbered column to be
subtracted from the well to the left of it in every odd column. Use
manual mapping to set up the appropriate map by placing the
standards, controls and samples in only the odd columns.
•
P-DOWN
enables a blank in the B, D, F and H rows to be subtracted
from the well above in the A, C, E and G rows. Use manual
mapping to set up the appropriate map by placing the standards,
controls, and samples in only the A, C, E, and G rows.
CONSTANT BLANK VALUE
This entry screen (Figure 3-28) only appears when a Constant Blank map is
selected. The user enters a value that is subtracted from each well on the
plate.
E N T E R B L A N K I N G
C O N S T A N T : 1 : 2 0
Figure 3-28:
Constant Blank Value screen
•
Use the NUMERIC keys to enter the value. The range is 0.000 to
3.000. The cursor is positioned at the first editable field and
advances automatically.
•
Use the CLEAR key to clear the value on the display.
Содержание PowerWave
Страница 1: ...PowerWaveTM Microplate Scanning Spectrophotometer Operator s Manual ...
Страница 2: ... ...
Страница 4: ......
Страница 6: ...ii ...
Страница 18: ...xiv ...
Страница 24: ...xx ...
Страница 91: ...PowerWave Operator s Manual 3 39 Figure 3 49A Panel Report ...
Страница 168: ...C 8 Error Codes ...
Страница 170: ...D 2 Report Format Figure 1 Samples with Calls on Matrix Report ...
Страница 171: ...Appendix D D 3 Figure 2 Curve Fit Report ...
Страница 172: ...D 4 Report Format Figure 3 Samples with Calls on Column Report ...
Страница 173: ...Appendix D D 5 Figure 4 Column Report without Samples ...
Страница 174: ...D 6 Report Format Figure 5 Panel Report ...
Страница 175: ...Appendix D D 7 Figure 6 Assay Detail Report ...
Страница 176: ...D 8 Report Format Figure 7 Assay List ...