C-4
Error
Codes
Displayed Error
Potential Cause
ERROR 0601
LAMBDA #1 Gain out of range
ERROR 0602
LAMBDA #2 Gain out of range
ERROR 0603
LAMBDA #3 Gain out of range
ERROR 0604
LAMBDA #4 Gain out of range
ERROR 0605
LAMBDA #5 Gain out of range
ERROR 0606
LAMBDA #6 Gain out of range
Errors 601-606 indicate that the gain for a specific wavelength is out of the
range necessary to ensure the filter’s performance to specifications.
P
ROBABLE CAUSE
:
•
A bad lamp could provide enough reduction in signal to cause
this error. Misaligned optics could also cause this error.
Displayed Error
Potential Cause
ERROR 0700
Reader failed noise test
Error 700 indicates that the measurement electronics do not read
repeatably at a maximum gain, with the light blocked.
P
ROBABLE CAUSE
:
•
External signals penetrating the measurement circuit. The
bottom and top shrouds, and the filter wheel cover should be
correctly installed. This problem in a correctly assembled unit
could indicate a problems with the reader’s printed circuit board.
Содержание PowerWave
Страница 1: ...PowerWaveTM Microplate Scanning Spectrophotometer Operator s Manual ...
Страница 2: ... ...
Страница 4: ......
Страница 6: ...ii ...
Страница 18: ...xiv ...
Страница 24: ...xx ...
Страница 91: ...PowerWave Operator s Manual 3 39 Figure 3 49A Panel Report ...
Страница 168: ...C 8 Error Codes ...
Страница 170: ...D 2 Report Format Figure 1 Samples with Calls on Matrix Report ...
Страница 171: ...Appendix D D 3 Figure 2 Curve Fit Report ...
Страница 172: ...D 4 Report Format Figure 3 Samples with Calls on Column Report ...
Страница 173: ...Appendix D D 5 Figure 4 Column Report without Samples ...
Страница 174: ...D 6 Report Format Figure 5 Panel Report ...
Страница 175: ...Appendix D D 7 Figure 6 Assay Detail Report ...
Страница 176: ...D 8 Report Format Figure 7 Assay List ...