B-22
Computer
Control
For each wavelength:
wavelength (3)
'200'-'800'
For each well selected (1 or 8):
comma separator (1)
‘,’
sign (1)
'+' or '-'
data (4)
‘xxxx'
well terminator
<CR> <LF>
data terminator
<^Z>
checksum (1)
0x00 - 0xFF
If the calibrate option is selected, all wavelengths indicated for the current spectral
scan will be calibrated. If calibration is not selected, any wavelengths selected but not
scanned since the instrument was powered up will be calibrated anyway. In
addition, any selected wavelengths with previous errors detected will also be
calibrated. If no calibration is performed at all, the <^Z> terminator is returned
alone with no wavelengths.
If a single well is selected for scanning, or if “last scan” is selected for a series option,
no first strip read is performed. In this case, the <^Z> terminator is returned alone
with no wavelengths.
Checksum calculation starts with the first byte AFTER the data start code, up through
and including the <^Z> data terminator code. The checksum is transmitted as one
hex data byte.
An aborted scan is indicated by a <DLE> response at any time before the final <^Z>
and checksum are received at the end of the response stream. Any data received,
including the checksum, should be considered invalid and irrelevant in the case of an
aborted scan.
NOTE: No assay name, report title, or time/date stamp - blank or otherwise - will
be returned with data collected using the above function.
Содержание PowerWave
Страница 1: ...PowerWaveTM Microplate Scanning Spectrophotometer Operator s Manual ...
Страница 2: ... ...
Страница 4: ......
Страница 6: ...ii ...
Страница 18: ...xiv ...
Страница 24: ...xx ...
Страница 91: ...PowerWave Operator s Manual 3 39 Figure 3 49A Panel Report ...
Страница 168: ...C 8 Error Codes ...
Страница 170: ...D 2 Report Format Figure 1 Samples with Calls on Matrix Report ...
Страница 171: ...Appendix D D 3 Figure 2 Curve Fit Report ...
Страница 172: ...D 4 Report Format Figure 3 Samples with Calls on Column Report ...
Страница 173: ...Appendix D D 5 Figure 4 Column Report without Samples ...
Страница 174: ...D 6 Report Format Figure 5 Panel Report ...
Страница 175: ...Appendix D D 7 Figure 6 Assay Detail Report ...
Страница 176: ...D 8 Report Format Figure 7 Assay List ...