B-20
Computer
Control
Each wavelength is represented by 'xxx,'. (3 ASCII digits followed by a comma).
Six wavelengths must be sent (24 bytes), followed by ETX after last comma.
If a wavelength is not to be specified, it must be replaced by '000'.
The
Set Filter Table
command returns the standard status response string.
After wavelength configuration, a self-test ('*' command) sequence must be run to
calibrate gains and generate self-check information for the new wavelength
configuration.
SPECTRAL SCANNING
The
Spectral Scan
('&') command is defined as follows:
Host Limits
Response
'&'
<ACK/NAK>
well (2)
'00': scan entire strip
'01' - '08': scan well in
indicated row
strip (2)
'01' - '12'
start wave (3)
'200' - '800'
stop wave (3)
'200' - '800'
wave step (3)
'1' - '600'
series option (1)
'0': single scan
‘1’:first scan in series
‘2’ : next scan
’3’ :last scan
Содержание PowerWave
Страница 1: ...PowerWaveTM Microplate Scanning Spectrophotometer Operator s Manual ...
Страница 2: ... ...
Страница 4: ......
Страница 6: ...ii ...
Страница 18: ...xiv ...
Страница 24: ...xx ...
Страница 91: ...PowerWave Operator s Manual 3 39 Figure 3 49A Panel Report ...
Страница 168: ...C 8 Error Codes ...
Страница 170: ...D 2 Report Format Figure 1 Samples with Calls on Matrix Report ...
Страница 171: ...Appendix D D 3 Figure 2 Curve Fit Report ...
Страница 172: ...D 4 Report Format Figure 3 Samples with Calls on Column Report ...
Страница 173: ...Appendix D D 5 Figure 4 Column Report without Samples ...
Страница 174: ...D 6 Report Format Figure 5 Panel Report ...
Страница 175: ...Appendix D D 7 Figure 6 Assay Detail Report ...
Страница 176: ...D 8 Report Format Figure 7 Assay List ...