B-18
Computer
Control
MULTIPLE WAVELENGTH READS
To cause a read to return data for each indicated wavelength (rather than performing
dual wavelength subtracts or blanking), use the following procedure:
a) Set both wavelengths in the standard reader definition table (see ‘V’
command) to '000'. In this case, the single/dual wavelength select parameter
(byte #30, bit #6) will be ignored.
b) Starting at byte #67 in the same table, replace blanking map values with six
wavelength's worth of ASCII characters (3 bytes each, 24 bytes total - if
wavelength not to be used, set to '000').
c) Send reader definition table to reader. ('V' command).
d) If reading a range of selected strip(s), send
Select Strips
command ('%') as
defined above.
e)
Send Read
command ('S') to perform the actual read.
Response data from the read will consist of an entire 8 x 12 plate for each indicated
wavelength, or an 8 x <#strips> plate consisting of strip(s) selected by the
Select
Strips
command (‘%’).
Wavelengths indicated by '000' will return no data.
NOTE:
Data returned for each plate read at an indicated wavelength will start with
<CR> followed by that wavelength’s well ODs as defined in the 312
protocol spec. Each plate will be terminated with a <^Z> character
followed by a checksum. Checksum calculation starts with the first byte
AFTER the data start code (<CR>), up through and including the <^Z>
data terminator code. The checksum is transmitted as a single hex data
byte.
Содержание PowerWave
Страница 1: ...PowerWaveTM Microplate Scanning Spectrophotometer Operator s Manual ...
Страница 2: ... ...
Страница 4: ......
Страница 6: ...ii ...
Страница 18: ...xiv ...
Страница 24: ...xx ...
Страница 91: ...PowerWave Operator s Manual 3 39 Figure 3 49A Panel Report ...
Страница 168: ...C 8 Error Codes ...
Страница 170: ...D 2 Report Format Figure 1 Samples with Calls on Matrix Report ...
Страница 171: ...Appendix D D 3 Figure 2 Curve Fit Report ...
Страница 172: ...D 4 Report Format Figure 3 Samples with Calls on Column Report ...
Страница 173: ...Appendix D D 5 Figure 4 Column Report without Samples ...
Страница 174: ...D 6 Report Format Figure 5 Panel Report ...
Страница 175: ...Appendix D D 7 Figure 6 Assay Detail Report ...
Страница 176: ...D 8 Report Format Figure 7 Assay List ...