Appendix C
C-3
Displayed Error
Potential Cause
ERROR 0400
Carrier axis failed position verify
ERROR 0401
Filter Wheel motor failed position
verify
ERROR 0402
Monochromator failed position
verify
ERROR 0403
Robot failed position verify
Errors 400, 401, 402 and 403 indicate that an axis failed its Position Verify
test. The Position Verify test monitors axis movements. If the axis moves
to its home position in too few or too many steps, the test fails.
P
ROBABLE CAUSE
:
•
The belt is slipping, due to incorrect tension, a loose motor pulley
or a loose belt clamp. This can also be caused by a bad motor
drive circuit.
Displayed Error
Potential Cause
ERROR 0500
Saturation error
Error 500 indicates an empty filter location on the filter wheel, or that the
filter wheel is not installed.
P
ROBABLE CAUSE
:
•
All wavelength locations must have either a bandpass or a filter
blank installed, or this error will occur. If the entire wheel is not
installed, the reader will also display this error.
Содержание PowerWave
Страница 1: ...PowerWaveTM Microplate Scanning Spectrophotometer Operator s Manual ...
Страница 2: ... ...
Страница 4: ......
Страница 6: ...ii ...
Страница 18: ...xiv ...
Страница 24: ...xx ...
Страница 91: ...PowerWave Operator s Manual 3 39 Figure 3 49A Panel Report ...
Страница 168: ...C 8 Error Codes ...
Страница 170: ...D 2 Report Format Figure 1 Samples with Calls on Matrix Report ...
Страница 171: ...Appendix D D 3 Figure 2 Curve Fit Report ...
Страница 172: ...D 4 Report Format Figure 3 Samples with Calls on Column Report ...
Страница 173: ...Appendix D D 5 Figure 4 Column Report without Samples ...
Страница 174: ...D 6 Report Format Figure 5 Panel Report ...
Страница 175: ...Appendix D D 7 Figure 6 Assay Detail Report ...
Страница 176: ...D 8 Report Format Figure 7 Assay List ...