Using the Detector
4
Operational Parameters
Light Source Intensity (LED)
The detector’s LED intensity can be adjusted in order to bring the
peak response back on-scale. The intensity range can be set between
1 and 100 %, with the default factory setting being 100 % for
maximum sensitivity. The LED setting is stored in memory and is
retained even after a power on/off cycle.
This feature is extremely useful for preparative chromatography
where samples of high concentration can be analyzed which would
otherwise exceed the dynamic range of the detector.
Power Mode
The detector can be configured from the front panel sub-menu (see
) to start in either RUN or STANDBY
mode when the detector is switched on via the rear power button.
To configure the power mode, select the required status mode (i.e.
STANDBY or RUN) you wish the detector to start-up in from the sub
menu screen (see
). The selected
option will take effect the next time the detector is power cycled.
If RUN mode is selected as the desired power mode, then the detector
will use the operating parameters stored in memory. In the unlikely
event that the detector encounters a fault during power-up, the
detector will automatically switch to STANDBY mode.
Data Output Rate (Hz)
The data output rate (Hz) setting is not used when using the detector
with an Isolera system.
NOTE
ELSD-A120 performs an automatic auto-zero (i.e. 10 mV) following an
LED change in order to keep the signal on-scale.
NOTE
All ELSD conditions are retained on power cycling of the detector.
Содержание ELSD-A120
Страница 1: ...Biotage ELSD A120 User Manual...
Страница 12: ...12 Biotage ELSD A120 User Manual 1 Introduction to the Detector Detector Overview...
Страница 20: ...20 Biotage ELSD A120 User Manual 2 Site Requirements and Specifications Performance Specifications...
Страница 44: ...44 Biotage ELSD A120 User Manual 5 Optimizing Performance Solvent Recommendations...
Страница 60: ...60 Biotage ELSD A120 User Manual 7 Error Information Error Messages...
Страница 74: ...74 Biotage ELSD A120 User Manual 8 Maintenance Replacing the Detector Firmware...
Страница 85: ......
Страница 86: ......