70
Biotage
®
ELSD-A120 User Manual
8
Maintenance
Replacing the Detector Firmware
Replacing the Detector Firmware
ELSD-A120 contains control firmware that can be flash upgraded via
the serial port.
To upgrade/downgrade the detectors firmware carry out the
following steps:
1
Connect the detector to a PC, via the serial port on the rear of the
detector, using the RS232 communication cable supplied with the
detector.
2
Open the Tera Term program, and select
File > New Connection
from the toolbar.
3
Select
Serial
and choose the appropriate COM port, followed by
OK
.
4
From the toolbar menu, select
Setup > Serial Port
to configure
the serial port connection.
When
For majority of internal repairs
Tools required
Description
Allen keys
Tera Term software
Current firmware file
Parts required
Description
Serial cable (supplied with the detector)
NOTE
Firmware versions v24.0.4 and v24.0.2, or later, are not compatible
with ELSD-A120.