Optimizing Performance
Mobile Phase Priming
Mobile Phase Priming
ELSD-A120 does not require any mobile phase priming, other than
that required to prime the solvent through the Isolera system. It is
recommended that priming of the Isolera system be performed
without ELSD-A120 attached, to prevent non-volatile impurities
contaminating the detector.
If operating in reversed-phase mode, the mobile phase should be
degassed and filtered, either by sparging with Helium or using
sonication.