7.  Program error messages 

7.1. 

Table  4  shows  the  program  cycle  error  messages,  their  description  and  possible 
solutions that can be made by user: 

Table 4. Program error messages 

Displayed message 

Description 

CANCELED BY OPERATOR 

Program execution is stopped by operator.  

Press 

Run/Stop

 key 

BOTTLE 4 OVERFLOWED 

Overflow of waste collection bottle. Empty the bottle. 

BOTTLE ## EMPTY 

Buffer bottle ## needs to be filled. Fill the bottle 

PLATE ERROR 

Plate calibration is required. See 

5.20

 

7.2. 

If the unit stops the operation and an unmentioned message appears, copy the error 
text, unpower the unit and contact your service company 

8.  Specifications 

The  unit  is  designed  for  operation  in  cold  rooms  and  closed  laboratory  rooms  at 

ambient temperature from +4°C to +40°C in a non-condensing atmosphere and maximum 
relative  humidity  80%  for  temperatures  up  to  31°C  decreasing  linearly  to  50%  relative 
humidity at 40°C. 

Biosan is committed to continuous impr

ovement of the unit’s properties and quality 

and reserves the right to alter unit’s design and specifications without additional notice. 

8.1. 

Dispense system of liquid dosage for each channel separately;  

8.2. 

Minimum dispense volume  ................................................................................ 

25 µl 

8.3. 

Maximum dispense volume  ........................................................................... 

1600 µl 

8.4. 

Dispense increment  .......................................................................................... 

25 µl 

8.5. 

Irregular liquid dosage at 300 µl ................................................ max. ±2.5% or 7.5 µl 

8.6. 

Allowed residual liquid volume in plate well ................................................ 

max. 2 µl 

8.7. 

Number of wells washed simultaneously ................................................................. 8 

8.8. 

Number of washing cycles ............................................................................... 1 - 15 

8.9. 

Aspiration time  ............................................................................................. 0.2 - 3 s 

8.10.  Aspiration/dispensing speed  .........................................................................3 levels 
8.11.  Choice of 3 washing buffers 

 

 

Содержание IW-8

Страница 1: ......

Страница 2: ...2...

Страница 3: ...Program error messages 16 8 Specifications 16 9 Ordering information 17 10 Maintenance and troubleshooting 18 11 Warranty 21 12 Glossary 22 13 EU Declaration of conformity 23 1 About this edition of...

Страница 4: ...rd the results obtained and repeat the washing procedure Accuracy of the obtained testing results depends on the quality of the microplate rinsing procedure To assure correctness of the rinsing proced...

Страница 5: ...erflow the maximum level during operation there is a mark on the side surface of the bottle Take necessary precautions utilizing waste liquid in accordance with general laboratory standards 2 4 AFTER...

Страница 6: ...W 8 the unit provides automatic control of reagent total volume in all bottles that is necessary for washing and waste collecting and estimates the volume of reagents necessary for washing Microbiolog...

Страница 7: ...bes outside inside diam length 6 3 600 mm 5 pieces 7 Tube for manifold outside inside diam length 3 2 1 6 400 mm 1 piece 8 Tube for manifold outside inside diam length 5 3 440 mm 1 piece 9 Tube for hy...

Страница 8: ...in a dry dust free place If the unit will be subjected to low temperatures dry the tubes as described in 6 Operation closedown Use the original packaging for transportation whether transported by air...

Страница 9: ...4 4 6 Install the manifold in the slots of the holder arm fig 2 3 Magnets in the manifold and the holder arm connect and align the manifold 4 4 7 Connect the manifold inlet connector supplying the li...

Страница 10: ...ce bottles near the unit 4 4 19 If 4 channel weight logger is included in the kit remove pads from underside of scale cups Place bottles on the 4 channel weight logger according to the numbering Conne...

Страница 11: ..._30 5 1x400_30 Soak 5 min Dispense 200 Dispense NO YES YES YES YES YES YES YES YES Aspirate YES YES YES YES YES YES YES YES YES Shake NO NO NO NO NO NO NO NO NO Dispense rate 02 02 02 02 02 02 02 02 0...

Страница 12: ...ttle to use for the main washing Second chan cycles 01 15 Number of washes with second solution 2 channel 1 3 Number of bottle to use for the second washing 5 4 Install the plate on the platform Atten...

Страница 13: ...y bottle number using and keys and press the Run Stop key 5 16 During operation display shows program number currently used bottle number and the number of cycles left fig 6 If the current action has...

Страница 14: ...section 6 Operation closedown of this manual Attention Always perform the rinsing cycle using distilled water after finishing the operation at the end of working day This will help to keep the unit in...

Страница 15: ...l figure 3 6 2 4 Press the Run Stop key fig 4 7 and the system performs a rinsing cycle Repeat the procedure twice if necessary 6 2 5 Repeat 6 2 2 6 2 4 for the remaining tubes 6 3 After finishing the...

Страница 16: ...ration in cold rooms and closed laboratory rooms at ambient temperature from 4 C to 40 C in a non condensing atmosphere and maximum relative humidity 80 for temperatures up to 31 C decreasing linearly...

Страница 17: ...IW 8 Intelispeed Washer V 4AW BS 060106 AAI 9 2 To inquire about or order the optional accessories or the replacement parts contact Biosan or your local Biosan representative 9 2 1 Optional accessori...

Страница 18: ...m the rinsing cycle twice using distilled water after finishing the operation see section 5 10 4 2 Monthly With 75 ethanol solution clean mounting surface of the plate platform and plate holder spring...

Страница 19: ...sed by needles they may contain dangerous infections Take off the manifold from the device put it down with needles upward Clean the aspiration channel needles with the wiping rod 0 80 mm and clean ne...

Страница 20: ...service engineer 10 8 Storage and transportation 10 8 1 Remove the manifold tubes and platform for plates 10 8 2 Put on packaging blocks onto sides of the unit and place it into the original carton bo...

Страница 21: ...ddress To obtain the claim form visit Technical support page on our website at link below 11 5 Extended warranty For IW 8 the Premium class model one year of extended warranty is available free of cha...

Страница 22: ...ich provides solution aspiration consequently from two opposite points of a well Strip washing mode plate washing by rows Dispensing filling wells with a set solution volume Waste bottle the bottle fo...

Страница 23: ...l equipment for measurement control and laboratory use General requirements EMC 2014 30 EU LVS EN 61326 1 2013 Electrical equipment for measurement control and laboratory use EMC requirements General...

Страница 24: ...Biosan SIA Ratsupites 7 k 2 Riga LV 1067 Latvia Phone 371 67426137 Fax 371 67428101 http www biosan lv Edition 4 02 November of 2020...

Отзывы: